UHPLC‐QTOFMS Urine Drug Screening With Dilute‐and‐Shoot Sample Preparation and Vacuum‐Insulated Probe‐Heated Electrospray Ionization

We developed a method for comprehensive urine drug screening by applying dilute-and-shoot extraction and vacuum-insulated probe-heated electrospray ionization with ultra-high performance liquid chromatography high-resolution quadrupole time-of-flight mass spectrometry (DS-UHPLC-VIP-HESI-QTOFMS). The method involved five-fold post-hydrolysis dilution of urine samples and chromatography on a C18 UHPLC column prior to QTOFMS analysis. The recently introduced VIP-HESI ion source was chosen due to its enhanced ionization efficiency and compatibility with UHPLC-QTOFMS. Extensive data was acquired in positive ion mode with a low collision energy (7 eV) and an elevated collision energy (30 eV), using the broadband collision-induced dissociation data acquisition scan mode that continuously generated high-resolution and accurate mass for parent and fragment qualifier ions, and parent ion isotopic patterns. Compound identification was performed against an in-house database with 1263 compound entries, using an automated post-run reverse target database search with preset identification criteria. Method validation with 56 different drugs showed acceptable results for the limit of identification (median 5 ng/mL), matrix effects (70-130%), repeatability of retention times (< 1%), mass accuracy (< 1 mDa), as well as for specificity and stability. As compared with an established UHPLC-QTOFMS method relying on solid-phase extraction and conventional electrospray ionization, DS-UHPLC-VIP-HESI-QTOFMS produced comparable results from authentic clinical urine samples for most drugs, but showed clearly improved detectability for pregabalin, gabapentin, and ritalinic acid. We anticipate that the new method will be a step forward for laboratories performing routine urine drug screening due to its fast turnaround time, reduced manual workload, cost efficiency, and broad substance coverage.

Authors

Institutions

Publication Details

Journal
Drug Testing and Analysis
Published
2024-11-12
DOI
https://doi.org/10.1002/dta.3830
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

UHPLC‐QTOFMS Urine Drug Screening With Dilute‐and‐Shoot Sample Preparation and Vacuum‐Insulated Probe‐Heated Electrospray Ionization

Carsten Baessmann, Anna Pelander, Mira Sundström, Pirkko Kriikku et al.
Drug Testing and Analysis
Mass Spectrometry Techniques and Applications
article

UHPLC‐QTOFMS Urine Drug Screening With Dilute‐and‐Shoot Sample Preparation and Vacuum‐Insulated Probe‐Heated Electrospray Ionization

Carsten Baessmann, Anna Pelander, Mira Sundström, Pirkko Kriikku, Ilkka Ojanperä
article en

Abstract

We developed a method for comprehensive urine drug screening by applying dilute-and-shoot extraction and vacuum-insulated probe-heated electrospray ionization with ultra-high performance liquid chromatography high-resolution quadrupole time-of-flight mass spectrometry (DS-UHPLC-VIP-HESI-QTOFMS). The method involved five-fold post-hydrolysis dilution of urine samples and chromatography on a C18 UHPLC column prior to QTOFMS analysis. The recently introduced VIP-HESI ion source was chosen due to its enhanced ionization efficiency and compatibility with UHPLC-QTOFMS. Extensive data was acquired in positive ion mode with a low collision energy (7 eV) and an elevated collision energy (30 eV), using the broadband collision-induced dissociation data acquisition scan mode that continuously generated high-resolution and accurate mass for parent and fragment qualifier ions, and parent ion isotopic patterns. Compound identification was performed against an in-house database with 1263 compound entries, using an automated post-run reverse target database search with preset identification criteria. Method validation with 56 different drugs showed acceptable results for the limit of identification (median 5 ng/mL), matrix effects (70-130%), repeatability of retention times (< 1%), mass accuracy (< 1 mDa), as well as for specificity and stability. As compared with an established UHPLC-QTOFMS method relying on solid-phase extraction and conventional electrospray ionization, DS-UHPLC-VIP-HESI-QTOFMS produced comparable results from authentic clinical urine samples for most drugs, but showed clearly improved detectability for pregabalin, gabapentin, and ritalinic acid. We anticipate that the new method will be a step forward for laboratories performing routine urine drug screening due to its fast turnaround time, reduced manual workload, cost efficiency, and broad substance coverage.

Drug Testing and AnalysisVol. 17(8)
University of Helsinki (FI), Knorr-Bremse (Germany) (DE), Finnish Institute for Health and Welfare (FI), Bruker (Germany) (DE)
Openalex Percentile: Top 85%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.