The thromboelastogram is confounded by hematocrit in clinical samples tested on both mechanical and acoustic platforms

Red blood cells are critical participants in normal hemostasis, but in vitro experiments have shown that the thromboelastogram (TEG) maximum amplitude has a paradoxical inverse relationship with hematocrit. This study reviewed all samples at a single academic institution where any mechanical (TEG 5000) or acoustic (TEG 6s) TEG was drawn within 5 min of hematocrit measurement. A total of 7,176 samples were identified using complete TEG and conventional coagulation test data (6,384 mechanical, 744 acoustic, and 48 both). In the primary analysis, hematocrit was negatively associated with the maximum amplitude on both mechanical and acoustic platforms after adjusting for relevant confounders. This suggests that the thromboelastogram may misrepresent the contribution of red blood cells in normal hemostasis.

Authors

Institutions

Publication Details

Journal
Frontiers in Medicine
Published
2024-11-13
DOI
https://doi.org/10.3389/fmed.2024.1421727
Primary Topic
Trauma, Hemostasis, Coagulopathy, Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The thromboelastogram is confounded by hematocrit in clinical samples tested on both mechanical and acoustic platforms

Aaron S. Hess
Frontiers in Medicine
Trauma, Hemostasis, Coagulopathy, Resuscitation
article

The thromboelastogram is confounded by hematocrit in clinical samples tested on both mechanical and acoustic platforms

Aaron S. Hess
article en

Abstract

Red blood cells are critical participants in normal hemostasis, but in vitro experiments have shown that the thromboelastogram (TEG) maximum amplitude has a paradoxical inverse relationship with hematocrit. This study reviewed all samples at a single academic institution where any mechanical (TEG 5000) or acoustic (TEG 6s) TEG was drawn within 5 min of hematocrit measurement. A total of 7,176 samples were identified using complete TEG and conventional coagulation test data (6,384 mechanical, 744 acoustic, and 48 both). In the primary analysis, hematocrit was negatively associated with the maximum amplitude on both mechanical and acoustic platforms after adjusting for relevant confounders. This suggests that the thromboelastogram may misrepresent the contribution of red blood cells in normal hemostasis.

Frontiers in MedicineVol. 11
University of Wisconsin–Madison (US)
Openalex Percentile: Top 74%
Trauma, Hemostasis, Coagulopathy, Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The thromboelastogram is confounded by hematocrit in clinical samples tested on both mechanical and acoustic platforms — Aaron S. Hess · Frontiers in Medicine (2024) | TGRS Research Map | TGRS