Abstract 4137564: Management of Atrial Flutter (AFL) Using Intrinsic Antitachycardia Pacing (iATP) Protocol

Background: The recently introduced iATP algorithm augments traditional ventricular antitachycardia pacing (ATP) with programmed stimulation (PS) resulting in greater efficacy terminating ventricular tachycardia than traditional ATP in avoiding ICD shocks. The use of iATP in the atrium may similarly enhance termination of atrial arrhythmias, avoiding prolonged arrhythmia burden or need for cardioversion. Objective: To compare efficacy of 3 atrial ATP protocols (PROT[GS1] ) in either converting AFL to sinus rhythm or accelerating AFL to atrial fibrillation (AF). Methods: Patients presenting in AFL for ablation were recruited. Patents underwent all three pacing protocols, twice, and in random order (6 times total). 1. PROT 1 – Ramp pacing at 91% of tachycardia cycle length (TCL) for 13 beats with 5 or 10ms decrement, avoiding pacing cycle length (CL) <160 ms. 2. PROT 2 – Ramp pacing at 81% of TCL for 13 beats again with 5 or 10ms decrement, avoiding CL <160 ms. 3. PROT 3 (iATP) – Drive train of 88% of TCL for 15 beats with singles, doubles, and triples down to atrial effective refractory period. If AFL terminated, it was re-induced with PS or burst pacing. Termination and CL acceleration to AF are defined as the positive effect. No effect is defined as the negative effect. CL acceleration and termination were treated similarly in the analysis because AF is clinically easier to manage than AFL. Results: Of the 21 patients enrolled, 17 completed the entire protocol. 4 patients (excluded from analysis) did not complete the protocol due to non-inducibility following AFL termination. The TCL of the AFL ranged from 190ms to 290ms. PROT 1 terminated AFL 6 out of 34 attempts (17.6%). PROT 2 terminated AFL 7 out of 34 attempts (20.5%). iATP terminated AFL 13 out of 34 attempts (38%). It also accelerated AFL more frequently than the other two protocols (PROT 1:1, 2:4, 3:8). Conclusion: Atrial iATP was significantly more effective in either terminating or accelerating AFL, compared to ramp pacing protocols.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4137564
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4137564: Management of Atrial Flutter (AFL) Using Intrinsic Antitachycardia Pacing (iATP) Protocol

Tharian S. Cherian, William A. Belden, George Shaw, Amit J. Thosani et al.
Circulation
Cardiac Arrhythmias and Treatments
article

Abstract 4137564: Management of Atrial Flutter (AFL) Using Intrinsic Antitachycardia Pacing (iATP) Protocol

Tharian S. Cherian, William A. Belden, George Shaw, Amit J. Thosani, Mati Friehling, Caitlin Phalunas, Joshua R. Silverstein, Manasvi Gupta, Emerson Liu, Minesh Lathia, Chaojing Duan, Elizabeth Belden, Grace Bailey
article en

Abstract

Background: The recently introduced iATP algorithm augments traditional ventricular antitachycardia pacing (ATP) with programmed stimulation (PS) resulting in greater efficacy terminating ventricular tachycardia than traditional ATP in avoiding ICD shocks. The use of iATP in the atrium may similarly enhance termination of atrial arrhythmias, avoiding prolonged arrhythmia burden or need for cardioversion. Objective: To compare efficacy of 3 atrial ATP protocols (PROT[GS1] ) in either converting AFL to sinus rhythm or accelerating AFL to atrial fibrillation (AF). Methods: Patients presenting in AFL for ablation were recruited. Patents underwent all three pacing protocols, twice, and in random order (6 times total). 1. PROT 1 – Ramp pacing at 91% of tachycardia cycle length (TCL) for 13 beats with 5 or 10ms decrement, avoiding pacing cycle length (CL) <160 ms. 2. PROT 2 – Ramp pacing at 81% of TCL for 13 beats again with 5 or 10ms decrement, avoiding CL <160 ms. 3. PROT 3 (iATP) – Drive train of 88% of TCL for 15 beats with singles, doubles, and triples down to atrial effective refractory period. If AFL terminated, it was re-induced with PS or burst pacing. Termination and CL acceleration to AF are defined as the positive effect. No effect is defined as the negative effect. CL acceleration and termination were treated similarly in the analysis because AF is clinically easier to manage than AFL. Results: Of the 21 patients enrolled, 17 completed the entire protocol. 4 patients (excluded from analysis) did not complete the protocol due to non-inducibility following AFL termination. The TCL of the AFL ranged from 190ms to 290ms. PROT 1 terminated AFL 6 out of 34 attempts (17.6%). PROT 2 terminated AFL 7 out of 34 attempts (20.5%). iATP terminated AFL 13 out of 34 attempts (38%). It also accelerated AFL more frequently than the other two protocols (PROT 1:1, 2:4, 3:8). Conclusion: Atrial iATP was significantly more effective in either terminating or accelerating AFL, compared to ramp pacing protocols.

CirculationVol. 150(Suppl_1)
Allegheny Health Network (US)
Good health and well-being
Openalex Percentile: Top 63%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.