Monte Carlo simulation of photo-peak efficiency response function and calculation of total efficiency for a NaI (Tl) scintillator detector

Abstract NaI (Tl) scintillation detectors are widely used in laboratories and in the field of gamma-ray spectrometry where detector efficiency is a key factor for successful quantitative γ ray analysis. In this work, Monte Carlo simulations using MCNPX code have been performed in order to evaluate energy resolution as a full width at half maximum (FWHM) of the peak and absolute efficiency of NaI detectors. Also, the detection efficiency and energy resolution for the NaI (Tl) scintillation detector of Tehran Research Reactor (TRR) pollution monitoring laboratory was measured. Simulation results were compared to experimental data and good agreement are obtained. In radioactive analyzing laboratory, precise determination of the detector efficiency according to any different gamma energy is a very important parameter whereas it uses in daily laboratory work.

Authors

Institutions

Publication Details

Journal
Kerntechnik
Published
2024-11-13
DOI
https://doi.org/10.1515/kern-2024-0093
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.61
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo simulation of photo-peak efficiency response function and calculation of total efficiency for a NaI (Tl) scintillator detector

Nahid Sadeghi, Zohreh Gholamzadeh, Mostafa Hassanzadeh
1 citations
Kerntechnik
Radiation Detection and Scintillator Technologies
0.61
article

Monte Carlo simulation of photo-peak efficiency response function and calculation of total efficiency for a NaI (Tl) scintillator detector

Nahid Sadeghi, Zohreh Gholamzadeh, Mostafa Hassanzadeh
article en
1 citations

Abstract

Abstract NaI (Tl) scintillation detectors are widely used in laboratories and in the field of gamma-ray spectrometry where detector efficiency is a key factor for successful quantitative γ ray analysis. In this work, Monte Carlo simulations using MCNPX code have been performed in order to evaluate energy resolution as a full width at half maximum (FWHM) of the peak and absolute efficiency of NaI detectors. Also, the detection efficiency and energy resolution for the NaI (Tl) scintillation detector of Tehran Research Reactor (TRR) pollution monitoring laboratory was measured. Simulation results were compared to experimental data and good agreement are obtained. In radioactive analyzing laboratory, precise determination of the detector efficiency according to any different gamma energy is a very important parameter whereas it uses in daily laboratory work.

KerntechnikVol. 89(6)
Atomic Energy Organization of Iran (IR), Nuclear Science and Technology Research Institute
Affordable and clean energy
Openalex Percentile: Top 29%
Radiation Detection and Scintillator Technologies
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.