A portable easy-to-use triboelectric sensor for arteriovenous fistula monitoring in dialysis patients

Hemodialysis is the primary treatment for patients with total loss of renal function and is performed through a vascular access called an arteriovenous fistula (AVF). AVF can be associated with several complications, such as stenosis, thrombosis, and aneurysms. While digital subtraction angiography (DSA) is the gold standard for AVF monitoring, its limited availability, invasiveness, and harmfulness has enabled Doppler ultrasound (DUS) as alternative technique. Herein, the development of a portable device based on a triboelectric sensor for AVF monitoring that exploits the concept of impedance cardiography (ICG) is presented. The triboelectric sensor consists of a thin polyvinylidene fluoride (PVDF) layer placed close to the AVF to evaluate the triboelectric ICG signal (T-ICG), thus avoiding the need for standard multiple-electrode equipment. Electrocardiographic (ECG) dry electrodes based on a composite of polydimethylsiloxane (PDMS) and multi-walled carbon nanotubes (MWCNTs) were fabricated. Concept validation was performed in a cohort of hemodialysis patients with and without AVF stenosis. The results demonstrated a linear dependence between blood flow and the maximum deviation (dZ/dt max ) of the T-ICG pattern. Patients with stenosis showed a statistically significant change in morphology at the B-point. The extent of stenosis was evaluated by analyzing the temporal slope of the ICG pattern in the B- and C-point regions, evidencing a linear trend for stenosis from 20 % up to 60 %. The proposed device represents an innovative approach compared with the gold clinical standards for the continuous monitoring of AVF in hemodialysis patients. • Non-invasive hemodynamic monitoring represents a key point in the development of clinical devices. • Triboelectric ICG pattern and synchronized ECG acquisition for a rapid and continuous patient monitoring. • Innovative approach for the continuous monitoring of AVF in hemodialysis patients. • Proof of concept of a portable triboelectric sensor to provide practical application oriented to clinical setting.

Authors

Institutions

Publication Details

Journal
Sensors International
Published
2024-11-13
DOI
https://doi.org/10.1016/j.sintl.2024.100309
Citations
4
Primary Topic
Advanced Sensor and Energy Harvesting Materials
Type
article
Field-Weighted Citation Impact
0.74
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A portable easy-to-use triboelectric sensor for arteriovenous fistula monitoring in dialysis patients

Antonino S. Fiorillo, Davide Bolignano, Pierangela Presta, Michele Andreucci et al.
4 citations
Sensors International
Advanced Sensor and Energy Harvesting Materials
0.74
article

A portable easy-to-use triboelectric sensor for arteriovenous fistula monitoring in dialysis patients

Antonino S. Fiorillo, Davide Bolignano, Pierangela Presta, Michele Andreucci, Giuseppe Coppolino, Michela Musolino, Salvatore A. Pullano, Nazareno Carullo, Giuseppe Oliva
article en
4 citations

Abstract

Hemodialysis is the primary treatment for patients with total loss of renal function and is performed through a vascular access called an arteriovenous fistula (AVF). AVF can be associated with several complications, such as stenosis, thrombosis, and aneurysms. While digital subtraction angiography (DSA) is the gold standard for AVF monitoring, its limited availability, invasiveness, and harmfulness has enabled Doppler ultrasound (DUS) as alternative technique. Herein, the development of a portable device based on a triboelectric sensor for AVF monitoring that exploits the concept of impedance cardiography (ICG) is presented. The triboelectric sensor consists of a thin polyvinylidene fluoride (PVDF) layer placed close to the AVF to evaluate the triboelectric ICG signal (T-ICG), thus avoiding the need for standard multiple-electrode equipment. Electrocardiographic (ECG) dry electrodes based on a composite of polydimethylsiloxane (PDMS) and multi-walled carbon nanotubes (MWCNTs) were fabricated. Concept validation was performed in a cohort of hemodialysis patients with and without AVF stenosis. The results demonstrated a linear dependence between blood flow and the maximum deviation (dZ/dt max ) of the T-ICG pattern. Patients with stenosis showed a statistically significant change in morphology at the B-point. The extent of stenosis was evaluated by analyzing the temporal slope of the ICG pattern in the B- and C-point regions, evidencing a linear trend for stenosis from 20 % up to 60 %. The proposed device represents an innovative approach compared with the gold clinical standards for the continuous monitoring of AVF in hemodialysis patients. • Non-invasive hemodynamic monitoring represents a key point in the development of clinical devices. • Triboelectric ICG pattern and synchronized ECG acquisition for a rapid and continuous patient monitoring. • Innovative approach for the continuous monitoring of AVF in hemodialysis patients. • Proof of concept of a portable triboelectric sensor to provide practical application oriented to clinical setting.

Sensors InternationalVol. 6
Magna Graecia University (IT)
Good health and well-being
Openalex Percentile: Top 31%
Advanced Sensor and Energy Harvesting Materials
0.74
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.