An Analysis of the Food and Drug Administration Manufacturer and User Facility Device Experience Database for MAGnetic Expansion Control Spinal Rods
These data represent the largest series of adverse events reported for MAGEC rods, including significant new data regarding MAGEC X. As well as endcap separation, failure of distraction, surface damage, and rod fracture were reported.
Authors
- Jack Filan
- Andrew Bowey
- TJ Joyce (ORCID: https://orcid.org/0000-0002-8497-5790)
Institutions
- Royal Victoria Infirmary (GB)
- Great North Children's Hospital (GB)
- Newcastle University (GB)
Publication Details
- Journal
- Therapeutic Innovation & Regulatory Science
- Published
- 2024-11-13
- DOI
- https://doi.org/10.1007/s43441-024-00724-4
- Citations
- 2
- Primary Topic
- Scoliosis diagnosis and treatment
- Type
- article
- Field-Weighted Citation Impact
- 1.15