Amplitude spectrum area is dependent on the electrocardiogram magnitude: evaluation of different normalization approaches

<i>Objective.</i>Amplitude Spectrum Area (AMSA) of the electrocardiogram (ECG) waveform during ventricular fibrillation (VF) has shown promise as a predictor of defibrillation success during cardiopulmonary resuscitation (CPR). However, AMSA relies on the magnitude of the ECG waveform, raising concerns about reproducibility across different settings that may introduce magnitude bias. This study aimed to evaluate different AMSA normalization approaches and their impact on removing bias while preserving predictive value.<i>Approach.</i>ECG were recorded in 118 piglets (1-2 months old) during a model of asphyxia-associated VF cardiac arrest and CPR. An initial subset (91/118) was recorded using one device (Device 1), and the remaining piglets were recorded in the second device (Device 2). Raw AMSA and three ECG magnitude metrics were estimated to assess magnitude-related bias between devices. Five AMSA normalization approaches were assessed for their ability to remove detected bias and to classify defibrillation success.<i>Main results.</i>Device 2 showed significantly lower ECG magnitude and raw AMSA compared to Device 1. CPR-based AMSA normalization approaches mitigated device-associated bias. Raw AMSA normalized by the average AMSA in the 1st minute of CPR (AMSA<sub>1m-cpr</sub>) exhibited the best sensitivity and specificity for classification of successful and unsuccessful defibrillation. While the optimal AMSA<sub>1m-cpr</sub>thresholds for balanced sensitivity and specificity were consistent across both devices, the optimal raw AMSA thresholds varied between the two devices. The area under the receiver operating characteristic curve for AMSA<sub>1m-cpr</sub>did not significantly differ from raw AMSA for both devices (Device 1: 0.74 vs. 0.88,<i>P</i>= 0.14; Device 2: 0.56 vs. 0.59,<i>P</i>= 0.81).<i>Significance.</i>Unlike raw AMSA, AMSA<sub>1m-cpr</sub>demonstrated consistent results across different devices while maintaining predictive value for defibrillation success. This consistency has important implications for the widespread use of AMSA and the development of future guidelines on optimal AMSA thresholds for successful defibrillation.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2024-11-01
DOI
https://doi.org/10.1088/1361-6579/ad9233
Citations
1
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.69

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Amplitude spectrum area is dependent on the electrocardiogram magnitude: evaluation of different normalization approaches

Nicholas J. Widmann, Fuchiang Tsui, Rodrigo M. Forti, Wesley B. Baker et al.
1 citations
Physiological Measurement
Cardiac Arrest and Resuscitation
0.69
article

Amplitude spectrum area is dependent on the electrocardiogram magnitude: evaluation of different normalization approaches

Nicholas J. Widmann, Fuchiang Tsui, Rodrigo M. Forti, Wesley B. Baker, Tiffany S. Ko, Hunter Gaudio, Ryan W. Morgan, Todd J. Kilbaugh, Viveknarayanan Padmanabhan, Kumaran Senthil, Luiz Eduardo Virgílio Silva, Constantine Mavroudis, Julia Slovis, Yuxi Lin, Lingyun Shi
article en
1 citations

Abstract

<i>Objective.</i>Amplitude Spectrum Area (AMSA) of the electrocardiogram (ECG) waveform during ventricular fibrillation (VF) has shown promise as a predictor of defibrillation success during cardiopulmonary resuscitation (CPR). However, AMSA relies on the magnitude of the ECG waveform, raising concerns about reproducibility across different settings that may introduce magnitude bias. This study aimed to evaluate different AMSA normalization approaches and their impact on removing bias while preserving predictive value.<i>Approach.</i>ECG were recorded in 118 piglets (1-2 months old) during a model of asphyxia-associated VF cardiac arrest and CPR. An initial subset (91/118) was recorded using one device (Device 1), and the remaining piglets were recorded in the second device (Device 2). Raw AMSA and three ECG magnitude metrics were estimated to assess magnitude-related bias between devices. Five AMSA normalization approaches were assessed for their ability to remove detected bias and to classify defibrillation success.<i>Main results.</i>Device 2 showed significantly lower ECG magnitude and raw AMSA compared to Device 1. CPR-based AMSA normalization approaches mitigated device-associated bias. Raw AMSA normalized by the average AMSA in the 1st minute of CPR (AMSA<sub>1m-cpr</sub>) exhibited the best sensitivity and specificity for classification of successful and unsuccessful defibrillation. While the optimal AMSA<sub>1m-cpr</sub>thresholds for balanced sensitivity and specificity were consistent across both devices, the optimal raw AMSA thresholds varied between the two devices. The area under the receiver operating characteristic curve for AMSA<sub>1m-cpr</sub>did not significantly differ from raw AMSA for both devices (Device 1: 0.74 vs. 0.88,<i>P</i>= 0.14; Device 2: 0.56 vs. 0.59,<i>P</i>= 0.81).<i>Significance.</i>Unlike raw AMSA, AMSA<sub>1m-cpr</sub>demonstrated consistent results across different devices while maintaining predictive value for defibrillation success. This consistency has important implications for the widespread use of AMSA and the development of future guidelines on optimal AMSA thresholds for successful defibrillation.

Physiological MeasurementVol. 45(11)
Children's Hospital of Philadelphia (US)
American Heart Association, Children's Hospital of Philadelphia, National Institutes of Health, National Heart, Lung, and Blood Institute, National Institute of Neurological Disorders and Stroke
Openalex Percentile: Top 24%
Cardiac Arrest and Resuscitation
0.69
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.