Assessment of Ultraviolet-C Light for Sterilization of Hysteroscopy Instruments

Objective To evaluate the sterilization efficacy of hysteroscopy instruments using ultraviolet C (UV-C) light at a wavelength of 259 nm in the Endoscopic Diagnostic Center of the National Institute of Perinatology. Methods Consecutive patients undergoing office hysteroscopy via the Bettochi vaginoscopy technique were included, excluding those with conditions such as viable intrauterine pregnancy or acute pelvic infection. Samples were collected from six designated sites of the hysteroscope, including the inner sheath, internal holes of the inner sheath, lens, graspers, scissors, and outer sheath. Initially, samples were taken after the first sterilization using a LAOKEN LK/MJG-150 Plasma Sterilizer (Chengdu, China). Next, samples were collected after the routine use of the hysteroscope in the office setting to confirm contamination. Subsequently, a new set of samples were taken after a 20-minute UV-C sterilization cycle with the EsteriUV device. Results The initial sterilization achieved a 96.73% sterilization rate, with ten samples testing positive for <i>Staphylococcus</i> <i>coagulase</i>-negative. Post-hysteroscopy, contamination increased significantly. Afterward, UV-C sterilization achieved a 96.08% sterilization rate, with 11 samples positive for <i>Staphylococcus</i> <i>coagulase</i>-negative and one for <i>Streptococcus anginosus</i> (p=0.66). No clinical infections were reported in any patient within one-month post-procedure. Conclusion UV-C light is a viable alternative for hysteroscopy instrument sterilization, demonstrating comparable efficacy to conventional methods. Further studies are recommended to optimize UV-C parameters for enhanced sterilization efficiency.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2024-11-13
DOI
https://doi.org/10.7759/cureus.73609
Citations
1
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.34
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Ultraviolet-C Light for Sterilization of Hysteroscopy Instruments

Luis Fernando Escobar-Ponce, Ricardo Figueroa-Damián, Jessica Aidée Mora-Galván, Andrea Olguín-Ortega et al.
1 citations
Cureus
Medical Device Sterilization and Disinfection
0.34
article

Assessment of Ultraviolet-C Light for Sterilization of Hysteroscopy Instruments

Luis Fernando Escobar-Ponce, Ricardo Figueroa-Damián, Jessica Aidée Mora-Galván, Andrea Olguín-Ortega, Graciela Villeda-Gabriel, Alejandro Rendón-Molina
article en
1 citations

Abstract

Objective To evaluate the sterilization efficacy of hysteroscopy instruments using ultraviolet C (UV-C) light at a wavelength of 259 nm in the Endoscopic Diagnostic Center of the National Institute of Perinatology. Methods Consecutive patients undergoing office hysteroscopy via the Bettochi vaginoscopy technique were included, excluding those with conditions such as viable intrauterine pregnancy or acute pelvic infection. Samples were collected from six designated sites of the hysteroscope, including the inner sheath, internal holes of the inner sheath, lens, graspers, scissors, and outer sheath. Initially, samples were taken after the first sterilization using a LAOKEN LK/MJG-150 Plasma Sterilizer (Chengdu, China). Next, samples were collected after the routine use of the hysteroscope in the office setting to confirm contamination. Subsequently, a new set of samples were taken after a 20-minute UV-C sterilization cycle with the EsteriUV device. Results The initial sterilization achieved a 96.73% sterilization rate, with ten samples testing positive for <i>Staphylococcus</i> <i>coagulase</i>-negative. Post-hysteroscopy, contamination increased significantly. Afterward, UV-C sterilization achieved a 96.08% sterilization rate, with 11 samples positive for <i>Staphylococcus</i> <i>coagulase</i>-negative and one for <i>Streptococcus anginosus</i> (p=0.66). No clinical infections were reported in any patient within one-month post-procedure. Conclusion UV-C light is a viable alternative for hysteroscopy instrument sterilization, demonstrating comparable efficacy to conventional methods. Further studies are recommended to optimize UV-C parameters for enhanced sterilization efficiency.

CureusVol. 16(11)
Instituto Nacional de Perinatología (MX)
Openalex Percentile: Top 35%
Medical Device Sterilization and Disinfection
0.34
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.