Abstract 4136286: Safety Events with a Large-Bore Aspiration Thrombectomy Device for Pulmonary Embolism: An analysis of the U.S. Food and Drug Administration Manufacturer and User Facility Device Experiences from 2018 to 2024

Introduction: Large bore thrombectomy is increasingly used to treat acute PE. Post-market surveillance using the U.S. FDA Manufacturer and User Facility Device Experience (MAUDE) database may capture serious device adverse events (AE) not described in small pre-market clinical trials. We aim to analyze serious AE associated with use of the Inari Medical (Irvine, California) large bore thrombectomy system to treat PE. Methods: All MAUDE events between January 1, 2018 and May 1, 2024 involving PE treatment with Triever and Flowtriever devices were reviewed and categorized based on device type, incident year, and clinical event. Based on narrative description, AE were placed into categories of cardiac, pulmonary, device malfunction, anemia, and other. Results: A total of 58 AE reports representing 50 unique clinical episodes were included. MAUDE event types were death (n=32), injury (n=23) and malfunction (n=3), and the majority (45/58) involved the larger aspiration catheters (Table 1). Death occurred in 27 of 50 unique episodes. Cardiac injury (n=19) and pulmonary injury (n=18) were most common and included 25 serious perforations. Serious AEs are listed in Table 2. Device malfunction/improper use was rare (n=7) and was not associated with death. Paradoxical embolism was described causing stroke and coronary occlusion. 3 of 4 blood loss AEs occurred prior to 2021. Conclusion: This report includes the largest description of reported AEs with large bore aspiration thrombectomy systems used to treat PE. This report describes serious events, often associated with death, that are likely infrequent and were not detected in pivotal studies or registries. Understanding these AEs may guide future technique and device refinement, and the current findings demonstrate the importance of large post-approval studies.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4136286
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4136286: Safety Events with a Large-Bore Aspiration Thrombectomy Device for Pulmonary Embolism: An analysis of the U.S. Food and Drug Administration Manufacturer and User Facility Device Experiences from 2018 to 2024

Logan Schwarzman, Nichelle Megowan, Shannon Lee, Joseph Thomas
Circulation
Venous Thromboembolism Diagnosis and Management
article

Abstract 4136286: Safety Events with a Large-Bore Aspiration Thrombectomy Device for Pulmonary Embolism: An analysis of the U.S. Food and Drug Administration Manufacturer and User Facility Device Experiences from 2018 to 2024

Logan Schwarzman, Nichelle Megowan, Shannon Lee, Joseph Thomas
article en

Abstract

Introduction: Large bore thrombectomy is increasingly used to treat acute PE. Post-market surveillance using the U.S. FDA Manufacturer and User Facility Device Experience (MAUDE) database may capture serious device adverse events (AE) not described in small pre-market clinical trials. We aim to analyze serious AE associated with use of the Inari Medical (Irvine, California) large bore thrombectomy system to treat PE. Methods: All MAUDE events between January 1, 2018 and May 1, 2024 involving PE treatment with Triever and Flowtriever devices were reviewed and categorized based on device type, incident year, and clinical event. Based on narrative description, AE were placed into categories of cardiac, pulmonary, device malfunction, anemia, and other. Results: A total of 58 AE reports representing 50 unique clinical episodes were included. MAUDE event types were death (n=32), injury (n=23) and malfunction (n=3), and the majority (45/58) involved the larger aspiration catheters (Table 1). Death occurred in 27 of 50 unique episodes. Cardiac injury (n=19) and pulmonary injury (n=18) were most common and included 25 serious perforations. Serious AEs are listed in Table 2. Device malfunction/improper use was rare (n=7) and was not associated with death. Paradoxical embolism was described causing stroke and coronary occlusion. 3 of 4 blood loss AEs occurred prior to 2021. Conclusion: This report includes the largest description of reported AEs with large bore aspiration thrombectomy systems used to treat PE. This report describes serious events, often associated with death, that are likely infrequent and were not detected in pivotal studies or registries. Understanding these AEs may guide future technique and device refinement, and the current findings demonstrate the importance of large post-approval studies.

CirculationVol. 150(Suppl_1)
Harbor–UCLA Medical Center (US)
Good health and well-being
Openalex Percentile: Top 60%
Venous Thromboembolism Diagnosis and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Abstract 4136286: Safety Events with a Large-Bore Aspiration Thrombectomy Device for Pulmonary Embolism: An analysis of the U.S. Food and Drug Administration Manufacturer and User Facility Device Experiences from 2018 to 2024 — Logan Schwarzman, Nichelle Megowan, et al. · Circulation (2024) | TGRS Research Map | TGRS