Tube current reduction and iterative image reconstruction for computed tomography myelography

This study aimed to systematically evaluate the impact of a low-dose (LD) protocol using tube current reduction on image quality, the confidence for intervention planning and guidance, and diagnostic yield for computed tomography (CT) myelography. We retrospectively analyzed 68 patients who underwent CT myelography, with 34 investigations performed with a standard-dose (SD) and 34 investigations performed with a LD protocol (using tube current reduction). The different scans were matched considering variables such as sex, age, presence of spinal instrumentation, and body diameter. All images were evaluated by two readers (R1 and R2) using Likert scales. Image noise was measured using attenuation values of paraspinal muscle tissue. Images were reconstructed with model-based iterative reconstruction (post-myelography diagnostic scans) or hybrid reconstruction (planning, periprocedural, and diagnostic scans). Image quality, overall artifacts, image contrast, and confidence for planning or intervention guidance were rated good to perfect for both SD and LD scans according to evaluations of both readers. Inter-reader agreement was good to very good for the images from intervention planning (κ ≥ 0.80) as well as for intervention guidance (κ ≥ 0.77), as well as for diagnostic scans (κ ≥ 0.85). Image noise was similar between SD and LD scans performed for planning of the interventional procedures (model-based iterative reconstruction: SD 45.37 ± 7.29 HU vs. LD 45.17 ± 9.12 HU; hybrid reconstruction: SD 46.05 ± 7.43 HU vs. LD 45.05 ± 8.69 HU; p > 0.05). The volume-weighted CT dose index (CTDI<sub>vol</sub>) and size-specific dose estimate (SSDE) were significantly lower for the planning scans as well as the periprocedural scans when using the LD protocol as compared to the SD protocol (p < 0.05). In conclusion, implementation of a LD protocol with tube current reduction for CT myelography is a feasible option to reduce radiation exposure, especially when combined with iterative image reconstruction. In our study, LD imaging did not have a relevant negative impact on image quality, confidence for intervention planning or guidance, or diagnostic certainty for CT myelography.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-11-12
DOI
https://doi.org/10.1038/s41598-024-79374-3
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.40

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Tube current reduction and iterative image reconstruction for computed tomography myelography

Nico Sollmann, Jan S. Kirschke, Vivian Schultz, Karina Kupfer et al.
1 citations
Scientific Reports
Radiation Dose and Imaging
0.40
article

Tube current reduction and iterative image reconstruction for computed tomography myelography

Nico Sollmann, Jan S. Kirschke, Vivian Schultz, Karina Kupfer, Meinrad Beer, Thomas Baum, Karolin J. Paprottka, Claus Zimmer
article en
1 citations

Abstract

This study aimed to systematically evaluate the impact of a low-dose (LD) protocol using tube current reduction on image quality, the confidence for intervention planning and guidance, and diagnostic yield for computed tomography (CT) myelography. We retrospectively analyzed 68 patients who underwent CT myelography, with 34 investigations performed with a standard-dose (SD) and 34 investigations performed with a LD protocol (using tube current reduction). The different scans were matched considering variables such as sex, age, presence of spinal instrumentation, and body diameter. All images were evaluated by two readers (R1 and R2) using Likert scales. Image noise was measured using attenuation values of paraspinal muscle tissue. Images were reconstructed with model-based iterative reconstruction (post-myelography diagnostic scans) or hybrid reconstruction (planning, periprocedural, and diagnostic scans). Image quality, overall artifacts, image contrast, and confidence for planning or intervention guidance were rated good to perfect for both SD and LD scans according to evaluations of both readers. Inter-reader agreement was good to very good for the images from intervention planning (κ ≥ 0.80) as well as for intervention guidance (κ ≥ 0.77), as well as for diagnostic scans (κ ≥ 0.85). Image noise was similar between SD and LD scans performed for planning of the interventional procedures (model-based iterative reconstruction: SD 45.37 ± 7.29 HU vs. LD 45.17 ± 9.12 HU; hybrid reconstruction: SD 46.05 ± 7.43 HU vs. LD 45.05 ± 8.69 HU; p > 0.05). The volume-weighted CT dose index (CTDI<sub>vol</sub>) and size-specific dose estimate (SSDE) were significantly lower for the planning scans as well as the periprocedural scans when using the LD protocol as compared to the SD protocol (p < 0.05). In conclusion, implementation of a LD protocol with tube current reduction for CT myelography is a feasible option to reduce radiation exposure, especially when combined with iterative image reconstruction. In our study, LD imaging did not have a relevant negative impact on image quality, confidence for intervention planning or guidance, or diagnostic certainty for CT myelography.

Scientific ReportsVol. 14(1)
TUM Klinikum (DE), University Hospital Ulm (DE), Technical University of Munich (DE)
Joachim Herz Stiftung, B. Braun-Stiftung, Leonhard-Lorenz-Stiftung, Technische Universität München
Good health and well-being
Openalex Percentile: Top 33%
Radiation Dose and Imaging
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.