Feasibility Study of Label‐Free Raman Spectroscopy for Parathyroid Gland Identification

We aim to evaluate the feasibility of Raman spectroscopy for parathyroid gland (PG) identification during thyroidectomy. Using a novel side-viewing handheld Raman probe, a total of 324 Raman spectra of four tissue types (i.e., thyroid, lymph node, PG, and lipid) commonly encountered during thyroidectomy were rapidly (< 3 s) acquired from 80 tissue sites (thyroid [n = 10], lymph node [n = 10], PG [n = 40], lipid [n = 20]) of 10 euthanized Wistar rats. Two partial least-squares (PLS)-discriminant analysis (DA) detection models were developed, differentiating the lipid and nonlipid (i.e., thyroid, lymph node, and PG) tissues with an accuracy of 100%, and PG, lymph node, and thyroid could be detected with an accuracy of 98.4%, 93.9%, and 95.4% respectively. This work demonstrates the feasibility of Raman spectroscopy technique for PG identification and protection during thyroidectomy at the molecular level.

Authors

Institutions

Publication Details

Journal
Journal of Biophotonics
Published
2024-11-11
DOI
https://doi.org/10.1002/jbio.202400220
Citations
1
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.66

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility Study of Label‐Free Raman Spectroscopy for Parathyroid Gland Identification

Jianfeng Wang, Ruoyu Zhang, Wen Tian, Hao Wang et al.
1 citations
Journal of Biophotonics
Spectroscopy Techniques in Biomedical and Chemical Research
0.66
article

Feasibility Study of Label‐Free Raman Spectroscopy for Parathyroid Gland Identification

Jianfeng Wang, Ruoyu Zhang, Wen Tian, Hao Wang, Sisi Guo, Jing Yao
article en
1 citations

Abstract

We aim to evaluate the feasibility of Raman spectroscopy for parathyroid gland (PG) identification during thyroidectomy. Using a novel side-viewing handheld Raman probe, a total of 324 Raman spectra of four tissue types (i.e., thyroid, lymph node, PG, and lipid) commonly encountered during thyroidectomy were rapidly (< 3 s) acquired from 80 tissue sites (thyroid [n = 10], lymph node [n = 10], PG [n = 40], lipid [n = 20]) of 10 euthanized Wistar rats. Two partial least-squares (PLS)-discriminant analysis (DA) detection models were developed, differentiating the lipid and nonlipid (i.e., thyroid, lymph node, and PG) tissues with an accuracy of 100%, and PG, lymph node, and thyroid could be detected with an accuracy of 98.4%, 93.9%, and 95.4% respectively. This work demonstrates the feasibility of Raman spectroscopy technique for PG identification and protection during thyroidectomy at the molecular level.

Journal of BiophotonicsVol. 18(1)
Beijing Institute of Technology (CN), Leibniz Institute of Photonic Technology (DE), Chinese PLA General Hospital (CN)
National Natural Science Foundation of China, Beijing Institute of Technology, Beijing Institute of Technology Research Fund Program for Young Scholars, Beijing Municipal Natural Science Foundation
Reduced inequalities
Openalex Percentile: Top 19%
Spectroscopy Techniques in Biomedical and Chemical Research
0.66
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Feasibility Study of Label‐Free Raman Spectroscopy for Parathyroid Gland Identification — Jianfeng Wang, Ruoyu Zhang, et al. · Journal of Biophotonics (2024) | TGRS Research Map | TGRS