Comparison of a Quantum Cascade Laser and an Interband Cascade Laser for the Detection of Stable Carbon Dioxide Isotopes Using Tunable Laser Absorption Spectroscopy

Quantum cascade lasers (QCLs) and interband cascade lasers (ICLs) are widely used as light sources in tunable laser absorption spectroscopy because they emit in the mid-infrared region where many strong and characteristic absorption bands are present. In this paper, we compare the performance of these lasers emitting at about 2310.1 cm<sup>-1</sup> to determine an optimal light source for detecting isotopic ratios of carbon dioxide (CO<sub>2</sub>). Our results show that the QCL has a higher relative intensity noise of up to 15 dBc/Hz compared to the ICL over the entire measured frequency range. In addition, it has a higher frequency fluctuation. However, the maximum tuning range of the QCL is up to 5.2 cm<sup>-1</sup> compared to up to 3.8 cm<sup>-1</sup> for the ICL. Both lasers lose more than half of their tuning range when the tuning rate is increased to 10 kHz. When measuring the isotope ratio of CO<sub>2</sub>, an uncertainty in the <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msup><mml:mi>δ</mml:mi><mml:mn>13</mml:mn></mml:msup></mml:math> value of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msubsup><mml:mi>σ</mml:mi><mml:mrow><mml:mrow><mml:mn>13</mml:mn><mml:mi>C</mml:mi><mml:mo>,</mml:mo><mml:mi>min</mml:mi></mml:mrow></mml:mrow><mml:mrow><mml:mrow><mml:mi>ICL</mml:mi></mml:mrow></mml:mrow></mml:msubsup><mml:mo>=</mml:mo><mml:mn>0.17</mml:mn></mml:math>‰ was achieved with the ICL and of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msubsup><mml:mi>σ</mml:mi><mml:mrow><mml:mrow><mml:mn>13</mml:mn><mml:mi>C</mml:mi><mml:mo>,</mml:mo><mml:mi>min</mml:mi></mml:mrow></mml:mrow><mml:mrow><mml:mrow><mml:mi>QCL</mml:mi></mml:mrow></mml:mrow></mml:msubsup><mml:mo>=</mml:mo><mml:mn>0.42</mml:mn></mml:math>‰ with the QCL, both at an integration time of 0.2 s. In summary, the QCL is more appropriate for applications that require a larger spectral tuning range, such as the measurement of a complex gas mixture, while the ICL has an excellent signal-to-noise ratio and is therefore better suited for applications that require higher precision.

Authors

Institutions

Publication Details

Journal
Applied Spectroscopy
Published
2024-11-11
DOI
https://doi.org/10.1177/00037028241291157
Citations
2
Primary Topic
Spectroscopy and Laser Applications
Type
article
Field-Weighted Citation Impact
0.48

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of a Quantum Cascade Laser and an Interband Cascade Laser for the Detection of Stable Carbon Dioxide Isotopes Using Tunable Laser Absorption Spectroscopy

Cem Dinc, Katrin Schmitt, P. Nitzsche, Leonard Nitzsche et al.
2 citations
Applied Spectroscopy
Spectroscopy and Laser Applications
0.48
article

Comparison of a Quantum Cascade Laser and an Interband Cascade Laser for the Detection of Stable Carbon Dioxide Isotopes Using Tunable Laser Absorption Spectroscopy

Cem Dinc, Katrin Schmitt, P. Nitzsche, Leonard Nitzsche, Jens Goldschmidt, Jürgen Wöllenstein
article en
2 citations

Abstract

Quantum cascade lasers (QCLs) and interband cascade lasers (ICLs) are widely used as light sources in tunable laser absorption spectroscopy because they emit in the mid-infrared region where many strong and characteristic absorption bands are present. In this paper, we compare the performance of these lasers emitting at about 2310.1 cm<sup>-1</sup> to determine an optimal light source for detecting isotopic ratios of carbon dioxide (CO<sub>2</sub>). Our results show that the QCL has a higher relative intensity noise of up to 15 dBc/Hz compared to the ICL over the entire measured frequency range. In addition, it has a higher frequency fluctuation. However, the maximum tuning range of the QCL is up to 5.2 cm<sup>-1</sup> compared to up to 3.8 cm<sup>-1</sup> for the ICL. Both lasers lose more than half of their tuning range when the tuning rate is increased to 10 kHz. When measuring the isotope ratio of CO<sub>2</sub>, an uncertainty in the <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msup><mml:mi>δ</mml:mi><mml:mn>13</mml:mn></mml:msup></mml:math> value of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msubsup><mml:mi>σ</mml:mi><mml:mrow><mml:mrow><mml:mn>13</mml:mn><mml:mi>C</mml:mi><mml:mo>,</mml:mo><mml:mi>min</mml:mi></mml:mrow></mml:mrow><mml:mrow><mml:mrow><mml:mi>ICL</mml:mi></mml:mrow></mml:mrow></mml:msubsup><mml:mo>=</mml:mo><mml:mn>0.17</mml:mn></mml:math>‰ was achieved with the ICL and of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:msubsup><mml:mi>σ</mml:mi><mml:mrow><mml:mrow><mml:mn>13</mml:mn><mml:mi>C</mml:mi><mml:mo>,</mml:mo><mml:mi>min</mml:mi></mml:mrow></mml:mrow><mml:mrow><mml:mrow><mml:mi>QCL</mml:mi></mml:mrow></mml:mrow></mml:msubsup><mml:mo>=</mml:mo><mml:mn>0.42</mml:mn></mml:math>‰ with the QCL, both at an integration time of 0.2 s. In summary, the QCL is more appropriate for applications that require a larger spectral tuning range, such as the measurement of a complex gas mixture, while the ICL has an excellent signal-to-noise ratio and is therefore better suited for applications that require higher precision.

Applied SpectroscopyVol. 79(5)
University of Freiburg (DE), Fraunhofer Institute for Physical Measurement Techniques (DE)
Deutsche Forschungsgemeinschaft
Clean water and sanitation
Openalex Percentile: Top 39%
Spectroscopy and Laser Applications
0.48
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.