Accuracy of the Apple Watch in Detecting Atrial Fibrillation Among Patients Undergoing 24-Hour Holter Monitoring: A Prospective, Pragmatic Study

BACKGROUND: As smartwatches with atrial fibrillation detection features gain popularity, it is important to assess the accuracy of these devices to guide decision-making. OBJECTIVES: Our study aimed to assess the sensitivity and specificity of the irregular rhythm notification and the electrocardiogram (ECG)–based detection features of a commonly used smart wearable device (Apple Watch) in detecting atrial fibrillation. METHODS: This was a prospective, pragmatic study conducted in Perpetual Succour Hospital–Cebu Heart Institute from August 2023 to January 2024. To assess the irregular rhythm notification feature, participants were asked to wear an Apple Watch alongside a 24-hour Holter monitor to verify notifications. For the ECG-based detection feature, participants had to tap the crown of the Apple Watch for 30 seconds to get a single-lead ECG similar to a lead I ECG tracing. They were instructed to get manual ECGs hourly, or more often while awake. Irregular rhythm notifications and ECG readings were then compared with that of the 24-hour Holter monitor. Sensitivity and specificity were then computed. RESULTS: A total of 140 participants consented to join after full study disclosure. The irregular rhythm notification feature of the Apple Watch exhibited a low sensitivity of 21.4% but achieved a high specificity of 100% in detecting atrial fibrillation. Meanwhile, the ECG-based detection feature, analyzed from 1295 manually taken ECGs with interpretable sinus rhythm or atrial fibrillation, demonstrated a high level of agreement with the Holter monitor, with a sensitivity of 100% and a specificity of 99.1%. CONCLUSION: The low sensitivity of the irregular rhythm notification feature of the Apple Watch in detecting atrial fibrillation cautions against relying on it as a primary screening tool. However, the high concordance of manually taken Apple Watch ECGs positions the device as a robust tool for detecting atrial fibrillation through manual ECG detection. KEYWORDS: Apple Watch, atrial fibrillation, smartwatches

Authors

Institutions

Publication Details

Journal
Philippine journal of cardiology.
Published
2024-11-11
DOI
https://doi.org/10.69944/pjc.4dfc7cfcda
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of the Apple Watch in Detecting Atrial Fibrillation Among Patients Undergoing 24-Hour Holter Monitoring: A Prospective, Pragmatic Study

Kazuhiro Katada, Elrey P. Inocian, Alex Junia, Mary Gertrude Ong Cordovez et al.
Philippine journal of cardiology.
Atrial Fibrillation Management and Outcomes
article

Accuracy of the Apple Watch in Detecting Atrial Fibrillation Among Patients Undergoing 24-Hour Holter Monitoring: A Prospective, Pragmatic Study

Kazuhiro Katada, Elrey P. Inocian, Alex Junia, Mary Gertrude Ong Cordovez, Louis Marie William Paday
article en

Abstract

BACKGROUND: As smartwatches with atrial fibrillation detection features gain popularity, it is important to assess the accuracy of these devices to guide decision-making. OBJECTIVES: Our study aimed to assess the sensitivity and specificity of the irregular rhythm notification and the electrocardiogram (ECG)–based detection features of a commonly used smart wearable device (Apple Watch) in detecting atrial fibrillation. METHODS: This was a prospective, pragmatic study conducted in Perpetual Succour Hospital–Cebu Heart Institute from August 2023 to January 2024. To assess the irregular rhythm notification feature, participants were asked to wear an Apple Watch alongside a 24-hour Holter monitor to verify notifications. For the ECG-based detection feature, participants had to tap the crown of the Apple Watch for 30 seconds to get a single-lead ECG similar to a lead I ECG tracing. They were instructed to get manual ECGs hourly, or more often while awake. Irregular rhythm notifications and ECG readings were then compared with that of the 24-hour Holter monitor. Sensitivity and specificity were then computed. RESULTS: A total of 140 participants consented to join after full study disclosure. The irregular rhythm notification feature of the Apple Watch exhibited a low sensitivity of 21.4% but achieved a high specificity of 100% in detecting atrial fibrillation. Meanwhile, the ECG-based detection feature, analyzed from 1295 manually taken ECGs with interpretable sinus rhythm or atrial fibrillation, demonstrated a high level of agreement with the Holter monitor, with a sensitivity of 100% and a specificity of 99.1%. CONCLUSION: The low sensitivity of the irregular rhythm notification feature of the Apple Watch in detecting atrial fibrillation cautions against relying on it as a primary screening tool. However, the high concordance of manually taken Apple Watch ECGs positions the device as a robust tool for detecting atrial fibrillation through manual ECG detection. KEYWORDS: Apple Watch, atrial fibrillation, smartwatches

Philippine journal of cardiology.Vol. 52(2)
Perpetual Succour Hospital (PH)
Peace, Justice and strong institutions
Openalex Percentile: Top 63%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.