Concurrent validity and accuracy of wrist-wearable devices to track heart rate during exercise in sedentary individuals

Background: Heart rate (HR) is commonly used as an indicator to represent a change in oxygen consumption and energy expenditure during activities. Nowadays, HR is easily measured by low-cost wrist-wearable devices, but a few studies examine the validity and accuracy of these devices in sedentary people. Objective: The study aimed to explore the concurrent validity and accuracy of low-cost wrist-wearable HR devices (The GT2e: approx. 4000 THB and Red Mi watch 2 lite: approx. 1000 THB) for measuring HR during exercise on a treadmill in people with a sedentary lifestyle. Materials and methods: Seventy-six sedentary participants (60 female; 78.95%) were instructed to wear a Polar H7 while randomly wearing a watch on each hand. Participants were asked to walk or run on the treadmill for 40 minutes, including free-living activities, exercise, and HR recovery phases. Pearson’s correlation coefficient was utilized to explain the levels of correlation (concurrent validity) of wrist wearable devices with the Polar H7. In contrast, the Bland-Altman method, Concordance correlation coefficient (CCC), and Mean Absolute Percentage Error (MAPE) were then used to determine the accuracy of wrist wearable devices. Results: The GT2e had an excellent agreement with the Polar H7 in free-living activities, exercise sessions, and the HR recovery phase (CCC=0.88, 0.85 & 0.78, respectively) and strong correlation with the criterion measured; Polar H7 (r=0.79 - 0.88; p<0.001). While Red Mi watch 2 lite also had an excellent agreement and correlation in free-living activities and the HR recovery phase (CCC=0.88 - 0.85; r=0.79), but a moderate agreement was found in the exercise phase (CCC=0.55; r=0.61 [p><0.001]). Conclusion: Both the GT2e and Red Mi watch 2 lite wrist-wearable devices could be used as alternative HR-measured devices to detect HR in sedentary people’s daily lives. However, the GT2e wrist-wearable device was more valid and accurate in detecting HR than the Red Mi watch 2 lite.

Authors

Publication Details

Journal
Journal of Associated Medical Sciences
Published
2024-11-11
DOI
https://doi.org/10.12982/jams.2025.017
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Concurrent validity and accuracy of wrist-wearable devices to track heart rate during exercise in sedentary individuals

Nithinun Chaikeeree, Rumpa Boonsinsukh, Teerawat Nithiatthawanon, Phatcharaphol Saenkap et al.
Journal of Associated Medical Sciences
Non-Invasive Vital Sign Monitoring
article

Concurrent validity and accuracy of wrist-wearable devices to track heart rate during exercise in sedentary individuals

Nithinun Chaikeeree, Rumpa Boonsinsukh, Teerawat Nithiatthawanon, Phatcharaphol Saenkap, Supawadi Arnthong, Janista Satheanpong, Apisara Kumjai, Praitawan Phantachang
article en

Abstract

Background: Heart rate (HR) is commonly used as an indicator to represent a change in oxygen consumption and energy expenditure during activities. Nowadays, HR is easily measured by low-cost wrist-wearable devices, but a few studies examine the validity and accuracy of these devices in sedentary people. Objective: The study aimed to explore the concurrent validity and accuracy of low-cost wrist-wearable HR devices (The GT2e: approx. 4000 THB and Red Mi watch 2 lite: approx. 1000 THB) for measuring HR during exercise on a treadmill in people with a sedentary lifestyle. Materials and methods: Seventy-six sedentary participants (60 female; 78.95%) were instructed to wear a Polar H7 while randomly wearing a watch on each hand. Participants were asked to walk or run on the treadmill for 40 minutes, including free-living activities, exercise, and HR recovery phases. Pearson’s correlation coefficient was utilized to explain the levels of correlation (concurrent validity) of wrist wearable devices with the Polar H7. In contrast, the Bland-Altman method, Concordance correlation coefficient (CCC), and Mean Absolute Percentage Error (MAPE) were then used to determine the accuracy of wrist wearable devices. Results: The GT2e had an excellent agreement with the Polar H7 in free-living activities, exercise sessions, and the HR recovery phase (CCC=0.88, 0.85 & 0.78, respectively) and strong correlation with the criterion measured; Polar H7 (r=0.79 - 0.88; p<0.001). While Red Mi watch 2 lite also had an excellent agreement and correlation in free-living activities and the HR recovery phase (CCC=0.88 - 0.85; r=0.79), but a moderate agreement was found in the exercise phase (CCC=0.55; r=0.61 [p><0.001]). Conclusion: Both the GT2e and Red Mi watch 2 lite wrist-wearable devices could be used as alternative HR-measured devices to detect HR in sedentary people’s daily lives. However, the GT2e wrist-wearable device was more valid and accurate in detecting HR than the Red Mi watch 2 lite.

Journal of Associated Medical SciencesVol. 58(1)
Affordable and clean energy
Openalex Percentile: Top 81%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.