Abstract Sa1107: Resuscitation of Lungs after Cardiac Arrest with Ex-Vivo Lung Perfusion: A Bench Study to Minimize Ventilator-Induced Lung Injury through Flow Modulation

Introduction: Lung donation after uncontrolled circulatory death (uDCD) following a witnessed cardiac arrest is feasible after resuscitation and assessment with ex vivo lung perfusion (EVLP). This bench study explores the efficacy of expiratory flow modulation in minimizing damage caused by mechanical power during volume-controlled ventilation within the EVLP procedure. Methods: Donor lungs were simulated using the ASL 5000 lung simulator (IngMar Medical) and ventilated in volume-controlled mode using the ViX-ventilator (KU Leuven), set at 7 breaths per minute with a positive end-expiratory pressure of 5 cmH 2 O. (Figure 1) Baseline measurements were obtained to deliver tidal volumes (VT) of 300, 400, and 500 mL. These measurements were then compared to an intervention where expiratory flow was modulated using an in-house designed device, the flow modulator. Mechanical power (J/min) was calculated using the pressure-volume curve. Differences in power between the baseline and various levels of flow modulation were assessed using a one-sided t-test. Results: No clinically relevant differences were observed in VT delivery between baseline settings and during expiratory flow modulation. (Table 1) However, significant differences in mechanical power were noted between baseline and flow modulation across the three VT settings (p<0.001). The reduction in power varied from 10.9% to 58.5%, depending on the selected VT and the degree of flow modulation applied. Notably, the extent of power reduction increased with the level of flow modulation. Conclusion: Flow modulation in the expiratory circuit during volume-controlled ventilation in a bench set-up does not affect VT delivery and can significantly reduce mechanical power by up to 58.5%. This reduction in mechanical power could potentially reduce iatrogenic damage during EVLP. Consequently, this might extend preservation times and improve the resuscitation outcomes of uncontrolled DCD lungs. Emergency medicine has the opportunity to leverage this vital pool of donor lungs, moving towards a future with minimal transplant waitlists.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-11
DOI
https://doi.org/10.1161/circ.150.suppl_1.sa1107
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract Sa1107: Resuscitation of Lungs after Cardiac Arrest with Ex-Vivo Lung Perfusion: A Bench Study to Minimize Ventilator-Induced Lung Injury through Flow Modulation

Jan Vercauteren, Benoît G. Marinus, Arne Neyrinck, Michiel Stiers et al.
Circulation
Cardiac Arrest and Resuscitation
article

Abstract Sa1107: Resuscitation of Lungs after Cardiac Arrest with Ex-Vivo Lung Perfusion: A Bench Study to Minimize Ventilator-Induced Lung Injury through Flow Modulation

Jan Vercauteren, Benoît G. Marinus, Arne Neyrinck, Michiel Stiers, Tom Schepens, Olivier Hoogmartens, Cornelia Genbrugge, Marc Sabbé, Joris Guldentops
article en

Abstract

Introduction: Lung donation after uncontrolled circulatory death (uDCD) following a witnessed cardiac arrest is feasible after resuscitation and assessment with ex vivo lung perfusion (EVLP). This bench study explores the efficacy of expiratory flow modulation in minimizing damage caused by mechanical power during volume-controlled ventilation within the EVLP procedure. Methods: Donor lungs were simulated using the ASL 5000 lung simulator (IngMar Medical) and ventilated in volume-controlled mode using the ViX-ventilator (KU Leuven), set at 7 breaths per minute with a positive end-expiratory pressure of 5 cmH 2 O. (Figure 1) Baseline measurements were obtained to deliver tidal volumes (VT) of 300, 400, and 500 mL. These measurements were then compared to an intervention where expiratory flow was modulated using an in-house designed device, the flow modulator. Mechanical power (J/min) was calculated using the pressure-volume curve. Differences in power between the baseline and various levels of flow modulation were assessed using a one-sided t-test. Results: No clinically relevant differences were observed in VT delivery between baseline settings and during expiratory flow modulation. (Table 1) However, significant differences in mechanical power were noted between baseline and flow modulation across the three VT settings (p<0.001). The reduction in power varied from 10.9% to 58.5%, depending on the selected VT and the degree of flow modulation applied. Notably, the extent of power reduction increased with the level of flow modulation. Conclusion: Flow modulation in the expiratory circuit during volume-controlled ventilation in a bench set-up does not affect VT delivery and can significantly reduce mechanical power by up to 58.5%. This reduction in mechanical power could potentially reduce iatrogenic damage during EVLP. Consequently, this might extend preservation times and improve the resuscitation outcomes of uncontrolled DCD lungs. Emergency medicine has the opportunity to leverage this vital pool of donor lungs, moving towards a future with minimal transplant waitlists.

CirculationVol. 150(Suppl_1)
Royal Military Academy (BE), KU Leuven (BE)
Good health and well-being
Openalex Percentile: Top 64%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.