Background: The Attain Stability Quad 4798 [Medtronic, Minneapolis, MN] lead was recently developed with a mechanism for active fixation into the venous wall, allowing for precise and secure placement in coronary venous branches. However, real-world data regarding complication rates are limited. Research Question: What are the real-world complications of the Medtronic 4798 lead? Goals: We reviewed the United States Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database for adverse events associated with the 4798 lead. Methods: We reviewed the MAUDE database on November 15, 2023, for all reported complications due to the 4798 lead from January 1, 2018 – a year before the lead was approved by the FDA – to the present. Results: A total of 2011 reports of adverse events occurred in mostly males (67.9%) patients with an average age of 70.86±12.1 years. Leads were used most frequently in 2023 (32.3%) for initial use (65.5%) but events commonly occurred intra-procedurally (73.7%). Device malfunction occurred in 43.4% of cases and included but was not limited to adverse events without an identified device problem (33.9%), device dislodgment (24.1%), infection (21.1%), and failure to capture (18.4%). In several cases, the guidewire was stuck (10.7%), or lead dislodgement (27.9%) occurred, resulting in direct helix damage (10.1%). Malfunctions resulted in injury and death in 55.8% and 1.1% of cases, respectively. Conclusions: The use of an active fixation CS lead has been associated with a significant incidence of intraprocedural helix mechanism malfunction. Further real-world surveillance and registry data can help clarify the true incidence of lead failure.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-11
DOI
https://doi.org/10.1161/circ.150.suppl_1.4146315
Primary Topic
Cardiac, Anesthesia and Surgical Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4146315: Complications Associated With the 4798 Attain Stability Active Fixation Coronary Sinus Lead: Insights from the MAUDE Registry

Dhanunjaya Lakkireddy, Donita Atkins, Naga Venkata K. Pothineni, Nikhil Gangasani et al.
Circulation
Cardiac, Anesthesia and Surgical Outcomes
article

Abstract 4146315: Complications Associated With the 4798 Attain Stability Active Fixation Coronary Sinus Lead: Insights from the MAUDE Registry

Dhanunjaya Lakkireddy, Donita Atkins, Naga Venkata K. Pothineni, Nikhil Gangasani, Aashish Katapadi, Rajesh Kabra, Nikhila Chelikam, Douglas Darden, Rakesh Gopinathannair, Aditya Mansabdar, Aashika Katapadi
article en

Abstract

Background: The Attain Stability Quad 4798 [Medtronic, Minneapolis, MN] lead was recently developed with a mechanism for active fixation into the venous wall, allowing for precise and secure placement in coronary venous branches. However, real-world data regarding complication rates are limited. Research Question: What are the real-world complications of the Medtronic 4798 lead? Goals: We reviewed the United States Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database for adverse events associated with the 4798 lead. Methods: We reviewed the MAUDE database on November 15, 2023, for all reported complications due to the 4798 lead from January 1, 2018 – a year before the lead was approved by the FDA – to the present. Results: A total of 2011 reports of adverse events occurred in mostly males (67.9%) patients with an average age of 70.86±12.1 years. Leads were used most frequently in 2023 (32.3%) for initial use (65.5%) but events commonly occurred intra-procedurally (73.7%). Device malfunction occurred in 43.4% of cases and included but was not limited to adverse events without an identified device problem (33.9%), device dislodgment (24.1%), infection (21.1%), and failure to capture (18.4%). In several cases, the guidewire was stuck (10.7%), or lead dislodgement (27.9%) occurred, resulting in direct helix damage (10.1%). Malfunctions resulted in injury and death in 55.8% and 1.1% of cases, respectively. Conclusions: The use of an active fixation CS lead has been associated with a significant incidence of intraprocedural helix mechanism malfunction. Further real-world surveillance and registry data can help clarify the true incidence of lead failure.

CirculationVol. 150(Suppl_1)
Heart Rhythm Society (US), University of Aruba (AW)
Good health and well-being
Openalex Percentile: Top 63%
Cardiac, Anesthesia and Surgical Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.