A Measurement Device for the Normal Spectral Emissivity of Materials Under Low-Temperature and Vacuum Conditions

Based on parallel light, a device for measuring sample emissivity under vacuum and low temperature was established. In this paper, a new emissivity measurement formula was designed, which replaces the derivation of Kirchhoff’s law of thermal radiation. The device is designed with a light-weight blackbody, an improved cooling speed, and an improved PID temperature control system to achieve a good temperature stability, close to 1 mK. The device is capable of measuring within a wavelength range of 4 μm to 16 μm. The results of the uncertainty assessment show that the uncertainty of the normal emissivity is better than 1.6% in the range of 4 μm to 7 μm, and better than 0.6% in the range of 7 μm to 14 μm (k = 1) at 0 °C.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2024-11-11
DOI
https://doi.org/10.3390/app142210369
Citations
1
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
1.30
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Measurement Device for the Normal Spectral Emissivity of Materials Under Low-Temperature and Vacuum Conditions

Guojin Feng, Xu‐Feng Liu, Zhenyuan Zhang
1 citations
Applied Sciences
Calibration and Measurement Techniques
1.30
article

A Measurement Device for the Normal Spectral Emissivity of Materials Under Low-Temperature and Vacuum Conditions

Guojin Feng, Xu‐Feng Liu, Zhenyuan Zhang
article en
1 citations

Abstract

Based on parallel light, a device for measuring sample emissivity under vacuum and low temperature was established. In this paper, a new emissivity measurement formula was designed, which replaces the derivation of Kirchhoff’s law of thermal radiation. The device is designed with a light-weight blackbody, an improved cooling speed, and an improved PID temperature control system to achieve a good temperature stability, close to 1 mK. The device is capable of measuring within a wavelength range of 4 μm to 16 μm. The results of the uncertainty assessment show that the uncertainty of the normal emissivity is better than 1.6% in the range of 4 μm to 7 μm, and better than 0.6% in the range of 7 μm to 14 μm (k = 1) at 0 °C.

Applied SciencesVol. 14(22)
National Institute of Metrology (CN), China Jiliang University (CN), Beijing University of Civil Engineering and Architecture (CN)
Affordable and clean energy
Openalex Percentile: Top 12%
Calibration and Measurement Techniques
1.30
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.