A statistical-based automatic detection of a low-contrast object in the ACR CT phantom for measuring contrast-to-noise ratio of CT images

<i>Purpose</i>. This study aimed to develop a new method for automated contrast-to-noise ratio (CNR) measurement using the low-contrast object in the ACR computed tomography (CT) phantom.<i>Methods</i>. The proposed method for CNR measurement was based on statistical criteria. A region of interest (ROI) was placed in a specific radial location and was then rotated around 360° in increments of 2°. At each position, the average CT number within the ROI was calculated. After one complete rotation, a profile of the average CT number around the full rotation was obtained. The center coordinate of the low-contrast object was determined from the maximum value of the profile. The CNR was calculated based on the average CT number and noise within the ROI in the low-contrast object and the ROI in the background, i.e., at the center of the phantom. The proposed method was used to evaluate CNR from images scanned with various phantom rotations, images with various noise levels (tube currents), and images from 25 CT scanners. The results were compared to a previous method based on a threshold approach.<i>Results</i>. The proposed method successfully placed the ROI properly in the center of a low-contrast object for variations of phantom rotation and tube current, whereas was not properly located in the center of the low-contrast object using the previous method. In addition, from 325 image samples of the 25 CT scanners, the proposed method successfully (100%) located the ROI within the low-contrast objects of all images used. The success rate of the previous method was only 58%.<i>Conclusion</i>. A new method for measuring CNR in the ACR CT phantom has been proposed and implemented. It is more powerful than a previous method based on a threshold approach.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-11-11
DOI
https://doi.org/10.1088/2057-1976/ad90e9
Citations
9
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
3.64
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A statistical-based automatic detection of a low-contrast object in the ACR CT phantom for measuring contrast-to-noise ratio of CT images

Ariij Naufal, Toshioh Fujibuchi, Choirul Anam, Geoff Dougherty et al.
9 citations
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
3.64
article

A statistical-based automatic detection of a low-contrast object in the ACR CT phantom for measuring contrast-to-noise ratio of CT images

Ariij Naufal, Toshioh Fujibuchi, Choirul Anam, Geoff Dougherty, Riska Amilia
article en
9 citations

Abstract

<i>Purpose</i>. This study aimed to develop a new method for automated contrast-to-noise ratio (CNR) measurement using the low-contrast object in the ACR computed tomography (CT) phantom.<i>Methods</i>. The proposed method for CNR measurement was based on statistical criteria. A region of interest (ROI) was placed in a specific radial location and was then rotated around 360° in increments of 2°. At each position, the average CT number within the ROI was calculated. After one complete rotation, a profile of the average CT number around the full rotation was obtained. The center coordinate of the low-contrast object was determined from the maximum value of the profile. The CNR was calculated based on the average CT number and noise within the ROI in the low-contrast object and the ROI in the background, i.e., at the center of the phantom. The proposed method was used to evaluate CNR from images scanned with various phantom rotations, images with various noise levels (tube currents), and images from 25 CT scanners. The results were compared to a previous method based on a threshold approach.<i>Results</i>. The proposed method successfully placed the ROI properly in the center of a low-contrast object for variations of phantom rotation and tube current, whereas was not properly located in the center of the low-contrast object using the previous method. In addition, from 325 image samples of the 25 CT scanners, the proposed method successfully (100%) located the ROI within the low-contrast objects of all images used. The success rate of the previous method was only 58%.<i>Conclusion</i>. A new method for measuring CNR in the ACR CT phantom has been proposed and implemented. It is more powerful than a previous method based on a threshold approach.

Biomedical Physics & Engineering ExpressVol. 11(1)
Diponegoro University (ID), Kyushu University (JP), California State University, Channel Islands (US)
Openalex Percentile: Top 6%
Medical Imaging Techniques and Applications
3.64
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.