PENERAPAN SIZE SPECIFIC DOSE ESTIMATE (SSDE) SEBAGAI TINGKAT PANDUAN DIAGNOSTIK LOKAL PADA PEMERIKSAAN CT SCAN KEPALA DI RS INDRIATI SOLOBARU, RSUD DR. MOEWARDI SURAKARTA, RSUD SLEMAN, DAN RS PUSAT OTAK NASIONAL JAKARTA
CT scan is a modality that produces a higher radiation dose compared to other modalities in radiodiagnostic installations. The radiation dose received by patients from examinations with this CT scan device is expressed in the volume CT Dose Index (CTDIvol) and Dose Length Product (DLP), which are dose indicators from the Indonesian Diagnostic Reference Level (DRL) value regulated in the Decree of the Head of the Nuclear Energy Regulatory Agency (BAPETEN) nationally. However, the CTDIvol value is only an indicator of the output dose from the CT scan, and the DLP value is an indicator of the total dose during the examination, which is influenced by the length of the scan. Both do not represent the specific dose received by the patient because the patient's dose depends not only on the output dose but also on the patient's characteristics. The patient's dose, which includes the output dose and patient characteristics, is called the Size-Specific Dose Estimate (SSDE), which is used to estimate the dose received by the patient by including a correction factor depending on the patient's dimensions. This study aims to calculate the SSDE value in a CT Head examination with a standard protocol using the same type of device, namely a 128-slice CT scan in several hospitals. The values of CTDIvol and DLP patients are obtained from the dose report displayed on the CT scan monitor screen in the IndoseCT software. The patient samples studied were patients undergoing CT Head examinations from March 2023 to February 2024, with 100 patients in each hospital. The results of this study obtained local DRL values based on SSDE calculated at Indriati Hospital, Moewardi Hospital, Sleman Hospital, and the National Brain Center Hospital and then compared them with the national TPD values. Based on the SSDE value from Indriati Hospital, the TPD value was 30.26 mGy, Sleman Hospital was 39.5 mGy, Dr. Moewardi Hospital was 49.3 mGy, and RSPON was 51.7 mGy. The TPD value of CT Head examinations without contrast from each hospital was still below the national TPD value of 60 mGy. Keywords: CT scan, diagnostic reference level (DRL), local DRL, national DRL, Size-Specific Dose Estimate (SSDE), IndoseCT.
Authors
- Dito Andi Rukmana
- Alviana Hapsari Purbarani
- Lutfiana Desy Saputri
- Adelina Oinike
Publication Details
- Journal
- Prosiding Seminar Si-INTAN
- Published
- 2024-11-06
- DOI
- https://doi.org/10.53862/ssi.v4.092024.015
- Primary Topic
- Management and Optimization Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00