OPTIMASI FAKTOR EKSPOSI DENGAN VARIASI KETEBALAN FILTER TERHADAP NILAI DOSIS DAN KUALITAS CITRA RADIOGRAFI PADA PEMERIKSAAN PELVIS

This study examines the impact of varying X-ray tube voltage (kV) and filter thickness on patient radiation dose and image quality in radiographic procedures. It aims to optimize exposure factors to minimize patient dose while maintaining diagnostic image quality, contributing to safer and more efficient radiological practices. This quantitative experimental study investigates the impact of varying filter thickness and tube voltage on patient radiation dose. Using X-ray equipment, dosimeters, and phantoms, measurements of Incident Air Kerma (INAK) and Entrance Surface Air Kerma (ESAK) and effective dose are taken, followed by image quality assessments to optimize exposure factors while maintaining diagnostic efficacy. The study demonstrated that increasing filter thickness decreases both INAK and ESAK values, with the most significant dose reduction occurring at 2 mm thickness. Optimal image quality and effective dose reduction were observed with the 1 mm filter at 80 kV, reducing the dose by 14.3% to 20% while maintaining good spatial resolution and contrast. The study concludes that using a 1 mm aluminum filter with 80 kV significantly reduces the effective dose while maintaining image quality. It is recommended that hospitals adopt additional filters, conduct further research with different materials, and optimize exposure factors. Training for radiology technicians on these practices is essential to enhance patient safety and image quality. Keywords: X-ray Equipment, Entrance Surface Air Kerma, Effective Dose, Image Quality.

Authors

Publication Details

Journal
Prosiding Seminar Si-INTAN
Published
2024-11-06
DOI
https://doi.org/10.53862/ssi.v4.092024.001
Primary Topic
Computer Science and Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

OPTIMASI FAKTOR EKSPOSI DENGAN VARIASI KETEBALAN FILTER TERHADAP NILAI DOSIS DAN KUALITAS CITRA RADIOGRAFI PADA PEMERIKSAAN PELVIS

Nooraishah Yasin, Muhammad Naufal Ramadhan, Raditya Faradina Pratiwi, Syalenda Rensapuan Maylana Fayza
Prosiding Seminar Si-INTAN
Computer Science and Engineering
article

OPTIMASI FAKTOR EKSPOSI DENGAN VARIASI KETEBALAN FILTER TERHADAP NILAI DOSIS DAN KUALITAS CITRA RADIOGRAFI PADA PEMERIKSAAN PELVIS

Nooraishah Yasin, Muhammad Naufal Ramadhan, Raditya Faradina Pratiwi, Syalenda Rensapuan Maylana Fayza
article en

Abstract

This study examines the impact of varying X-ray tube voltage (kV) and filter thickness on patient radiation dose and image quality in radiographic procedures. It aims to optimize exposure factors to minimize patient dose while maintaining diagnostic image quality, contributing to safer and more efficient radiological practices. This quantitative experimental study investigates the impact of varying filter thickness and tube voltage on patient radiation dose. Using X-ray equipment, dosimeters, and phantoms, measurements of Incident Air Kerma (INAK) and Entrance Surface Air Kerma (ESAK) and effective dose are taken, followed by image quality assessments to optimize exposure factors while maintaining diagnostic efficacy. The study demonstrated that increasing filter thickness decreases both INAK and ESAK values, with the most significant dose reduction occurring at 2 mm thickness. Optimal image quality and effective dose reduction were observed with the 1 mm filter at 80 kV, reducing the dose by 14.3% to 20% while maintaining good spatial resolution and contrast. The study concludes that using a 1 mm aluminum filter with 80 kV significantly reduces the effective dose while maintaining image quality. It is recommended that hospitals adopt additional filters, conduct further research with different materials, and optimize exposure factors. Training for radiology technicians on these practices is essential to enhance patient safety and image quality. Keywords: X-ray Equipment, Entrance Surface Air Kerma, Effective Dose, Image Quality.

Prosiding Seminar Si-INTANVol. 4(1)
Openalex Percentile: Top 82%
Computer Science and Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.