Radiation dose assessment of pediatric computed tomography of the chest: the need to consider patient size
CTDIvol systematically underestimated radiation dose in comparison to SSDE in pediatric patients submitted to chest CT and should not be used as the primary parameter to monitor CT protocols in these patients. SSDE calculated according to effective diameter also underestimates the radiation dose compared to SSDE calculated based on water equivalent.