Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards

The uMI Panorama GS PET/CT system is a new long-axial-field-of-view scanner featuring high sensitivity, time-of-flight (TOF) resolution, spatial resolution, and count rate performance. The aim of this study is to assess the PET system on the basis of the National Electrical Manufacturers Association (NEMA) NU 2-2018 and European Association of Nuclear Medicine Research Limited (EARL) standards. <b>Methods:</b> Spatial resolution, count rate performance, sensitivity, accuracy, image quality, TOF resolution, and coregistration accuracy were evaluated following the NEMA NU 2-2018 standard. Additional experiments included energy resolution, 200-cm-long line sources for sensitivity, a 175-cm-long scatter phantom for count rate and TOF resolution, as well as the compliance with the EARL guideline. Moreover, an <sup>18</sup>F-FDG PET patient study was reconstructed with various frame durations. <b>Results:</b> The PET system achieved sub-3-mm transaxial and axial spatial resolutions at a 1-cm radial offset. The sensitivities with the 70-cm-long and 200-cm-long line sources were observed to be 176.3 and 90.8 kcps/MBq, respectively, at the center of the field of view. The noise-equivalent count rates (NECRs) of the 70-cm-long and 175-cm-long scatter phantoms were measured to be 3.35 Mcps at 57.57 kBq/mL and 2.24 Mcps at 33.27 kBq/mL, respectively. The TOF resolutions for both phantoms were approximately 189 ps at 5.3 kBq/mL and lower than 200 ps below the NECR peaks. The absolute count rate errors of all 34 acquisitions were less than 3% below the NECR peak for the 70-cm-long scatter phantom. With the standard NEMA image quality phantom experiment, the contrast recovery coefficient varied from 68.17% to 94.20% and the background variabilities were all below 2%. The maximum PET/CT coregistration error was 1.33 mm. Regarding EARL compliance, the gaussian filter of 5-mm full width at half maximum could produce acceptable images. The patient data demonstrate visually satisfactory image quality with short frames (less than 1 min). <b>Conclusion:</b> The uMI Panorama GS exhibits spatial resolution and TOF resolution similar to those of the uMI Panorama system (35-cm axial field of view), despite the extended axial field of view. The 148-cm axial coverage, sub-200-ps TOF resolution, high sensitivity, and count rate performances are expected to yield superior image quality and offer new opportunities for various clinical applications.

Authors

Institutions

Publication Details

Journal
Journal of Nuclear Medicine
Published
2024-11-07
DOI
https://doi.org/10.2967/jnumed.124.267963
Citations
15
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
6.06
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards

Yu Liu, Meixi Liu, Haiqun Xing, Li Huo et al.
15 citations
Journal of Nuclear Medicine
Medical Imaging Techniques and Applications
6.06
article

Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards

Yu Liu, Meixi Liu, Haiqun Xing, Li Huo, Chao Ren, Xinchun Yan, Haiqiong Zhang, Zhixin Hao
article en
15 citations

Abstract

The uMI Panorama GS PET/CT system is a new long-axial-field-of-view scanner featuring high sensitivity, time-of-flight (TOF) resolution, spatial resolution, and count rate performance. The aim of this study is to assess the PET system on the basis of the National Electrical Manufacturers Association (NEMA) NU 2-2018 and European Association of Nuclear Medicine Research Limited (EARL) standards. <b>Methods:</b> Spatial resolution, count rate performance, sensitivity, accuracy, image quality, TOF resolution, and coregistration accuracy were evaluated following the NEMA NU 2-2018 standard. Additional experiments included energy resolution, 200-cm-long line sources for sensitivity, a 175-cm-long scatter phantom for count rate and TOF resolution, as well as the compliance with the EARL guideline. Moreover, an <sup>18</sup>F-FDG PET patient study was reconstructed with various frame durations. <b>Results:</b> The PET system achieved sub-3-mm transaxial and axial spatial resolutions at a 1-cm radial offset. The sensitivities with the 70-cm-long and 200-cm-long line sources were observed to be 176.3 and 90.8 kcps/MBq, respectively, at the center of the field of view. The noise-equivalent count rates (NECRs) of the 70-cm-long and 175-cm-long scatter phantoms were measured to be 3.35 Mcps at 57.57 kBq/mL and 2.24 Mcps at 33.27 kBq/mL, respectively. The TOF resolutions for both phantoms were approximately 189 ps at 5.3 kBq/mL and lower than 200 ps below the NECR peaks. The absolute count rate errors of all 34 acquisitions were less than 3% below the NECR peak for the 70-cm-long scatter phantom. With the standard NEMA image quality phantom experiment, the contrast recovery coefficient varied from 68.17% to 94.20% and the background variabilities were all below 2%. The maximum PET/CT coregistration error was 1.33 mm. Regarding EARL compliance, the gaussian filter of 5-mm full width at half maximum could produce acceptable images. The patient data demonstrate visually satisfactory image quality with short frames (less than 1 min). <b>Conclusion:</b> The uMI Panorama GS exhibits spatial resolution and TOF resolution similar to those of the uMI Panorama system (35-cm axial field of view), despite the extended axial field of view. The 148-cm axial coverage, sub-200-ps TOF resolution, high sensitivity, and count rate performances are expected to yield superior image quality and offer new opportunities for various clinical applications.

Journal of Nuclear MedicineVol. 65(12)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Peking Union Medical College Hospital (CN)
Affordable and clean energy
Openalex Percentile: Top 3%
Medical Imaging Techniques and Applications
6.06
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.