Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields
In all investigated algorithms, error increases in the penumbra areas, outside the field, and at cross-plane of open fields and in-plane direction of wedged fields. By increasing the wedge angle, the discrepancy between the TPS algorithms and MC simulations becomes more pronounced. This discrepancy is attributed to the increased presence of scattered photons and the variation in the delivered dose within the wedge field, consequently impacts the beam quality. While the CCC and FFT algorithms had better accuracy, the Clarkson algorithm, particularly at larger effective wedge angles, exhibited greater effectiveness than the two mentioned algorithms.