Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields

In all investigated algorithms, error increases in the penumbra areas, outside the field, and at cross-plane of open fields and in-plane direction of wedged fields. By increasing the wedge angle, the discrepancy between the TPS algorithms and MC simulations becomes more pronounced. This discrepancy is attributed to the increased presence of scattered photons and the variation in the delivered dose within the wedge field, consequently impacts the beam quality. While the CCC and FFT algorithms had better accuracy, the Clarkson algorithm, particularly at larger effective wedge angles, exhibited greater effectiveness than the two mentioned algorithms.

Authors

Institutions

Publication Details

Journal
Journal of Medical Signals & Sensors
Published
2024-11-01
DOI
https://doi.org/10.4103/jmss.jmss_28_24
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields

Mohammad Ehsan Ravari, Marziyeh Behmadi, Hamid-Reza Sadoughi, Sajjad Raghavi
1 citations
Journal of Medical Signals & Sensors
Advanced Radiotherapy Techniques
0.61
article

Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields

Mohammad Ehsan Ravari, Marziyeh Behmadi, Hamid-Reza Sadoughi, Sajjad Raghavi
article en
1 citations

Abstract

In all investigated algorithms, error increases in the penumbra areas, outside the field, and at cross-plane of open fields and in-plane direction of wedged fields. By increasing the wedge angle, the discrepancy between the TPS algorithms and MC simulations becomes more pronounced. This discrepancy is attributed to the increased presence of scattered photons and the variation in the delivered dose within the wedge field, consequently impacts the beam quality. While the CCC and FFT algorithms had better accuracy, the Clarkson algorithm, particularly at larger effective wedge angles, exhibited greater effectiveness than the two mentioned algorithms.

Journal of Medical Signals & SensorsVol. 14(11)
Semnan University of Medical Sciences (IR), North Khorasan University of Medical Sciences (IR)
Semnan University
Good health and well-being
Openalex Percentile: Top 29%
Advanced Radiotherapy Techniques
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.