Commissioning of a novel gantry-less proton therapy system

Purpose The focus of this article is to describe the configuration, testing, and commissioning of a novel gantry-less synchrotron-based proton therapy (PT) facility. Materials and methods The described PT system delivers protons with a water equivalent range between 4 and 38 cm in 1800 energy layers. The fixed beam delivery permits a maximum field size of 28 × 30 cm 2 . The patient positioning and imaging system includes a six-degree-of-freedom robotic arm, a convertible patient chair, a vertical 4DCT, and an orthogonal 2D X-ray imaging system. Results The spot positioning reproducibility was consistent within ±1 mm. The width (σ) of the beam profile at the isocenter was energy dependent and ranged from 2.8 mm to 7.7 mm. Absolute dose reproducibility was measured and deviations were found to be <0.62% for all possible beam scenarios. The built-in dose monitoring system was successfully tested for its ability to generate interlocks under specific conditions (beam spot deviation ≥2 mm, individual spot dose ≥10% or ≥0.25 Gy, spot energy deviation ≥0.5 MeV). The robot positioning exhibited a consistent reproducibility within ±1 mm. All tested scenarios achieved laser-free initial 3D/3D image-guided positioning within ±5 mm. Subsequent 2D/3D positioning showed an accuracy of ±1 mm. A single 2D/3D image registration event corrected positions in all cases. Results of gamma analysis (3%, 3 mm) demonstrated pass rates greater than 95% for head and neck, thorax, abdomen treatment plans. Conclusions We report on the performance of a novel single-room gantry-less PT system comprised of a compact synchrotron and an adjustable (from nearly horizontal to almost vertical) patient positioning system. The commissioning results show high accuracy and reproducibility of the main proton beam parameters and the patient positioning system. The new PT facility started patient treatments in March 2023, which were the first in Israel and the Middle Eastern region.

Authors

Institutions

Publication Details

Journal
Frontiers in Oncology
Published
2024-11-07
DOI
https://doi.org/10.3389/fonc.2024.1417393
Citations
7
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
2.93
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Commissioning of a novel gantry-less proton therapy system

Aron Popovtzer, Yair Hillman, Alexander Pryanichnikov, Jon Feldman et al.
7 citations
Frontiers in Oncology
Radiation Therapy and Dosimetry
2.93
article

Commissioning of a novel gantry-less proton therapy system

Aron Popovtzer, Yair Hillman, Alexander Pryanichnikov, Jon Feldman, Philip Blumenfeld, Alejandro Achkienasi, Ilya Polyansky, Michael Marash, Stas Raskin
article en
7 citations

Abstract

Purpose The focus of this article is to describe the configuration, testing, and commissioning of a novel gantry-less synchrotron-based proton therapy (PT) facility. Materials and methods The described PT system delivers protons with a water equivalent range between 4 and 38 cm in 1800 energy layers. The fixed beam delivery permits a maximum field size of 28 × 30 cm 2 . The patient positioning and imaging system includes a six-degree-of-freedom robotic arm, a convertible patient chair, a vertical 4DCT, and an orthogonal 2D X-ray imaging system. Results The spot positioning reproducibility was consistent within ±1 mm. The width (σ) of the beam profile at the isocenter was energy dependent and ranged from 2.8 mm to 7.7 mm. Absolute dose reproducibility was measured and deviations were found to be <0.62% for all possible beam scenarios. The built-in dose monitoring system was successfully tested for its ability to generate interlocks under specific conditions (beam spot deviation ≥2 mm, individual spot dose ≥10% or ≥0.25 Gy, spot energy deviation ≥0.5 MeV). The robot positioning exhibited a consistent reproducibility within ±1 mm. All tested scenarios achieved laser-free initial 3D/3D image-guided positioning within ±5 mm. Subsequent 2D/3D positioning showed an accuracy of ±1 mm. A single 2D/3D image registration event corrected positions in all cases. Results of gamma analysis (3%, 3 mm) demonstrated pass rates greater than 95% for head and neck, thorax, abdomen treatment plans. Conclusions We report on the performance of a novel single-room gantry-less PT system comprised of a compact synchrotron and an adjustable (from nearly horizontal to almost vertical) patient positioning system. The commissioning results show high accuracy and reproducibility of the main proton beam parameters and the patient positioning system. The new PT facility started patient treatments in March 2023, which were the first in Israel and the Middle Eastern region.

Frontiers in OncologyVol. 14
German Cancer Research Center (DE), Hebrew University of Jerusalem (IL), Heidelberg University (DE), Hadassah Medical Center (IL)
Affordable and clean energy
Openalex Percentile: Top 8%
Radiation Therapy and Dosimetry
2.93
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.