ANALYSIS OF UNCERTAINTY SOURCES IN DOSE MEASUREMENT AT AN INDUSTRIAL ELECTRON ACCELERATOR

The procedure for in-situ/in-plant calibration of routine dosimeters used to support the product processing at the LU-10 industrial electron linac of NSC KIPT is described. The standard polystyrene calorimeters are used as reference dosimeters. The uncertainty budget for dose measurements with the Harwell Red Perspex 4034 routine dosimeters has been compiled. Based on the analysis of numerous measurements of absorbed dose using polystyrene calorimeters, the changes in their metrological characteristics have been observed related both to the duration of their use and the amount of absorbed dose accumulated over that period. The data obtained confirm the need for recalibrating reference dosimeters by an accredited calibration lab, as well as the importance of participating in the dosimetry intercomparisons, including those conducted under the IAEA projects.

Authors

Publication Details

Published
2024-10-25
DOI
https://doi.org/10.46813/2024-153-117
Citations
1
Primary Topic
Graphite, nuclear technology, radiation studies
Type
article
Field-Weighted Citation Impact
0.12

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ANALYSIS OF UNCERTAINTY SOURCES IN DOSE MEASUREMENT AT AN INDUSTRIAL ELECTRON ACCELERATOR

V.L. Uvarov, D.V. Titov, V.A. Shevchenko, S.K. Romanovsky et al.
1 citations
Graphite, nuclear technology, radiation studies
0.12
article

ANALYSIS OF UNCERTAINTY SOURCES IN DOSE MEASUREMENT AT AN INDUSTRIAL ELECTRON ACCELERATOR

V.L. Uvarov, D.V. Titov, V.A. Shevchenko, S.K. Romanovsky, V.Yu. Titov, R. Pomatsalyuk
article en
1 citations

Abstract

The procedure for in-situ/in-plant calibration of routine dosimeters used to support the product processing at the LU-10 industrial electron linac of NSC KIPT is described. The standard polystyrene calorimeters are used as reference dosimeters. The uncertainty budget for dose measurements with the Harwell Red Perspex 4034 routine dosimeters has been compiled. Based on the analysis of numerous measurements of absorbed dose using polystyrene calorimeters, the changes in their metrological characteristics have been observed related both to the duration of their use and the amount of absorbed dose accumulated over that period. The data obtained confirm the need for recalibrating reference dosimeters by an accredited calibration lab, as well as the importance of participating in the dosimetry intercomparisons, including those conducted under the IAEA projects.

International Atomic Energy Agency
Industry, innovation and infrastructure
Openalex Percentile: Top 58%
Graphite, nuclear technology, radiation studies
0.12
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

ANALYSIS OF UNCERTAINTY SOURCES IN DOSE MEASUREMENT AT AN INDUSTRIAL ELECTRON ACCELERATOR — V.L. Uvarov, D.V. Titov, et al. · (2024) | TGRS Research Map | TGRS