In this study, the imaging performance of electronic portal imaging devices (EPIDs) is evaluated, comparing measurements collected from EPID images captured at 115 cm, with a field size of 15 × 15 cm2, monitor units (MUs) in the range of 2 MU-100 MU and dose rates (DRs) of 200 MU/min, 400 MU/min and 600 MU/min, using a 6 MV LINAC system and the QC-3V image quality phantom. The analysis includes the normalized contrast transfer function (CTFnorm), the noise power spectrum (NPS) and the information capacity (IC), as well as the signal-to-noise frequency response (SNFR), which can be used as a comprehensive quality index. The results of our study are compared with previously published data captured at 100 cm under similar exposure conditions. They show similar CTF curves with different source-to-phantom distances, with the lowest values observed at specific MU and DR combinations. Moreover, NPS graphs are found to decrease with respect to spatial frequency. SNFR values also display a reduction with increasing spatial frequency. In addition, irradiation with the phantom placed closer to the EPID, 115 cm from the LINAC, yields better SNFR and IC performance characteristics, indicating better delineation of the organs closer to the EPID. The testing of EPID performance may potentially benefit from our results, which may lead to improvements in the quality of radiotherapy treatments.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2024-11-07
DOI
https://doi.org/10.3390/app142210260
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Image Quality and Information Parameters of Electronic Portal Imaging Devices

Vasiliki Peppa, Ioannis Valais, Nektarios Kalyvas, Georgios Karakatsanis et al.
1 citations
Applied Sciences
Medical Imaging Techniques and Applications
0.40
article

Image Quality and Information Parameters of Electronic Portal Imaging Devices

Vasiliki Peppa, Ioannis Valais, Nektarios Kalyvas, Georgios Karakatsanis, C. Michail, George Fountos, Marios K. Tzomakas, Antigoni Alexiou, Anastasios Episkopakis, I. Kandarakis
article en
1 citations

Abstract

In this study, the imaging performance of electronic portal imaging devices (EPIDs) is evaluated, comparing measurements collected from EPID images captured at 115 cm, with a field size of 15 × 15 cm2, monitor units (MUs) in the range of 2 MU-100 MU and dose rates (DRs) of 200 MU/min, 400 MU/min and 600 MU/min, using a 6 MV LINAC system and the QC-3V image quality phantom. The analysis includes the normalized contrast transfer function (CTFnorm), the noise power spectrum (NPS) and the information capacity (IC), as well as the signal-to-noise frequency response (SNFR), which can be used as a comprehensive quality index. The results of our study are compared with previously published data captured at 100 cm under similar exposure conditions. They show similar CTF curves with different source-to-phantom distances, with the lowest values observed at specific MU and DR combinations. Moreover, NPS graphs are found to decrease with respect to spatial frequency. SNFR values also display a reduction with increasing spatial frequency. In addition, irradiation with the phantom placed closer to the EPID, 115 cm from the LINAC, yields better SNFR and IC performance characteristics, indicating better delineation of the organs closer to the EPID. The testing of EPID performance may potentially benefit from our results, which may lead to improvements in the quality of radiotherapy treatments.

Applied SciencesVol. 14(22)
National and Kapodistrian University of Athens (GR), University of West Attica (GR), Alexandra Hospital (GR)
Openalex Percentile: Top 33%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.