The study aims to evaluate the energy response of the thermoluminescent dosemeter (TLD) CaSO4:Dy from high energy photon beams produced from medical linear accelerator. The test was performed on the polymethyl methacrylate phantom surface and at the depth of dose maximum with a source-to-surface distance of 100 cm and the radiation field size of 12 × 12 cm. The results were compared with the TLD response exposed to 60Co standard source. The results show that the TLD response to the linac's photon differs >30% compared to the TLD response to 60Co exposure when measured at the phantom surface, while at the depth of maximum dose the response is similar (< ±10%), indicating that the TLD response is no longer dependent on the energy. This implies that the current dose reading evaluation based on 60Co calibration curve should be corrected to obtain more accurate dose report of the radiation workers.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2024-11-07
DOI
https://doi.org/10.1093/rpd/ncae211
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Energy response of CaSO4:Dy thermoluminescence detector to medical linear accelerator photon beams

Bunawas Bunawas, Sri Herwiningsih, Lia Wilda Izzati
Radiation Protection Dosimetry
Advanced X-ray and CT Imaging
article

Energy response of CaSO4:Dy thermoluminescence detector to medical linear accelerator photon beams

Bunawas Bunawas, Sri Herwiningsih, Lia Wilda Izzati
article en

Abstract

The study aims to evaluate the energy response of the thermoluminescent dosemeter (TLD) CaSO4:Dy from high energy photon beams produced from medical linear accelerator. The test was performed on the polymethyl methacrylate phantom surface and at the depth of dose maximum with a source-to-surface distance of 100 cm and the radiation field size of 12 × 12 cm. The results were compared with the TLD response exposed to 60Co standard source. The results show that the TLD response to the linac's photon differs >30% compared to the TLD response to 60Co exposure when measured at the phantom surface, while at the depth of maximum dose the response is similar (< ±10%), indicating that the TLD response is no longer dependent on the energy. This implies that the current dose reading evaluation based on 60Co calibration curve should be corrected to obtain more accurate dose report of the radiation workers.

Radiation Protection DosimetryVol. 200(20)
University of Brawijaya (ID)
Affordable and clean energy
Openalex Percentile: Top 82%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.