Troponin Test, Not Only a Number: An Unusual Case of False Positive

Heterophile antibodies, which can arise from infections, autoimmune disorders, or exposure to animal antigens, can interfere with immunoassays. These antibodies can cross-react with the test reagents used in troponin assays, causing a false elevation in troponin levels. The paper describes a case of a 37-year-old male drug abuser admitted to the emergency room with chest pain. A series of troponin measurements performed using different assays gave discrepant results. Only thanks to the use of Scantibodies HBT tubes, which remove heterophile antibodies, was it possible to make a correct diagnosis of troponin negativity. In conclusion, a correct laboratory/clinical approach to the identification of heterophile antibody interference is essential for accurate troponin testing in order to avoid false positive results. Implementing neutralizing tests can significantly improve the reliability of these diagnostic assays, ensuring better patient outcome.

Authors

Institutions

Publication Details

Journal
International Journal of Molecular Sciences
Published
2024-11-06
DOI
https://doi.org/10.3390/ijms252211937
Citations
2
Primary Topic
Acute Myocardial Infarction Research
Type
article
Field-Weighted Citation Impact
0.80

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Troponin Test, Not Only a Number: An Unusual Case of False Positive

Michela Salvatici, Francesca Rispoli, Lorenzo Drago, Francesco Bandera et al.
2 citations
International Journal of Molecular Sciences
Acute Myocardial Infarction Research
0.80
article

Troponin Test, Not Only a Number: An Unusual Case of False Positive

Michela Salvatici, Francesca Rispoli, Lorenzo Drago, Francesco Bandera, Barbara Bianchi, Eleonora Matteucci, Andrea Antonelli, Monica Gaimarri, Delia Francesca Sansico
article en
2 citations

Abstract

Heterophile antibodies, which can arise from infections, autoimmune disorders, or exposure to animal antigens, can interfere with immunoassays. These antibodies can cross-react with the test reagents used in troponin assays, causing a false elevation in troponin levels. The paper describes a case of a 37-year-old male drug abuser admitted to the emergency room with chest pain. A series of troponin measurements performed using different assays gave discrepant results. Only thanks to the use of Scantibodies HBT tubes, which remove heterophile antibodies, was it possible to make a correct diagnosis of troponin negativity. In conclusion, a correct laboratory/clinical approach to the identification of heterophile antibody interference is essential for accurate troponin testing in order to avoid false positive results. Implementing neutralizing tests can significantly improve the reliability of these diagnostic assays, ensuring better patient outcome.

International Journal of Molecular SciencesVol. 25(22)
University of Milan (IT), MultiMedica (IT)
Ministero della Salute
Good health and well-being
Openalex Percentile: Top 23%
Acute Myocardial Infarction Research
0.80
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.