Differences in power consumption of continuous positive airway pressure (CPAP) devices

<bold>Objective:</bold> CPAP is the most common treatment for obstructive sleep apnea but it's electrical consumption (EC) is unknown despite its economic and ecological impact. Our aim was to compare the EC of 6 CPAP devices and to assess the impact of a humidification system (HS). <bold>Methods:</bold> Each device (A to F) was bench tested at pressures from 4 to 20 cmH2O, without and with HS. Real-life data from 40818 patients were extrapolated at the national level to compare the potential impact on EC and related costs of using the most energy-efficient device versus the most energy-consuming one (using 200€/ Gigawatt-hour (GWH)). <bold>Results:</bold> Compared with the device with the lower EC (device B), EC varied from +6% to +40% depending on the device (p<0.001). EC increased linearly with pressure level. The addition of a HS increased the EC with an average increase of +345% (+250% to +395% (p<0.001)). Patients from the cohort were treated by 75% of A, 5% of B, 3% of C, 1% of D, 15% of E, and 1% of F with a mean compliance of 6.9 hours, a mean pressure of 7 cmH2O for one year. Fitting the entire French population with the least energy-intensive CPAP device (B) without HS would result in savings of 111680 GWH and 22M€ per year compared to a widespread of the most energy-intensive CPAP (C) with HS. <bold>Conclusion:</bold> The choice of CPAP, its settings and the addition of a HS have a major impact on EC, on the cost and ultimately on the carbon footprint linked to EC. <fig><object-id>erj;64/suppl_68/OA5437/F1</object-id><object-id>F1</object-id><object-id>F1</object-id><graphic></graphic></fig>

Authors

Institutions

Publication Details

Published
2024-09-14
DOI
https://doi.org/10.1183/13993003.congress-2024.oa5437
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Differences in power consumption of continuous positive airway pressure (CPAP) devices

Arnaud Prigent, Clara Bianquis, Maxime Patout, Emeline Fresnel et al.
Respiratory Support and Mechanisms
article

Differences in power consumption of continuous positive airway pressure (CPAP) devices

Arnaud Prigent, Clara Bianquis, Maxime Patout, Emeline Fresnel, Isabelle Arnulf, Adrien Kerfourn, J. Dupuis, Wojciech Trzepizur, Frédéric Gagnadoux, Emilie Cottet
article en

Abstract

<bold>Objective:</bold> CPAP is the most common treatment for obstructive sleep apnea but it's electrical consumption (EC) is unknown despite its economic and ecological impact. Our aim was to compare the EC of 6 CPAP devices and to assess the impact of a humidification system (HS). <bold>Methods:</bold> Each device (A to F) was bench tested at pressures from 4 to 20 cmH2O, without and with HS. Real-life data from 40818 patients were extrapolated at the national level to compare the potential impact on EC and related costs of using the most energy-efficient device versus the most energy-consuming one (using 200€/ Gigawatt-hour (GWH)). <bold>Results:</bold> Compared with the device with the lower EC (device B), EC varied from +6% to +40% depending on the device (p<0.001). EC increased linearly with pressure level. The addition of a HS increased the EC with an average increase of +345% (+250% to +395% (p<0.001)). Patients from the cohort were treated by 75% of A, 5% of B, 3% of C, 1% of D, 15% of E, and 1% of F with a mean compliance of 6.9 hours, a mean pressure of 7 cmH2O for one year. Fitting the entire French population with the least energy-intensive CPAP device (B) without HS would result in savings of 111680 GWH and 22M€ per year compared to a widespread of the most energy-intensive CPAP (C) with HS. <bold>Conclusion:</bold> The choice of CPAP, its settings and the addition of a HS have a major impact on EC, on the cost and ultimately on the carbon footprint linked to EC. <fig><object-id>erj;64/suppl_68/OA5437/F1</object-id><object-id>F1</object-id><object-id>F1</object-id><graphic></graphic></fig>

Délégation Paris 5 (FR), Inserm (FR), Université Sorbonne Nouvelle (FR), Sorbonne Université (FR), Santé Publique France (FR), Assistance Publique – Hôpitaux de Paris (FR), Saint Laurent Polyclinique (FR), Pitié-Salpêtrière Hospital (FR), Centre Hospitalier Universitaire de Rouen (FR), Université d'Angers (FR)
Affordable and clean energy
Openalex Percentile: Top 63%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.