Performance of low-flow oxygen delivery systems with mouth breathing and altered respiratory mechanics: a bench study.
<bold>BACKGROUND:</bold> Scientific gaps exist regarding the efficacy and optimal use of low-flow oxygen delivery systems in various clinical scenarios. The aim of this study was to compare the FiO<sub>2</sub> measured by different low-flow systems in breathing conditions with opened or closed mouth, and in a normal, restrictive or obstructive respiratory pattern. <bold>Methods:</bold> An active artificial lung (ASL 5000) was connected to a three-dimensional printed head sequentially equipped with nasal canula (NC), simple facemask (FM), non-rebreathing facemask (NRM), NC + surgical facemask (NC_Sfm) or NC + double-trunk mask (NC_Dtm). Oxygen flow supplied NC, NC_Sfm and NC_Dtm from 1 to 10L/min, FM from 5 to 10L/min and NRM at 10L/min. The experiments were carried out with three lung mechanics (normal, obstructive and restrictive) with and without mouth closed. The FiO<sub>2</sub> sampled in the artificial lung was measured. <bold>Results:</bold> From 1 to 4 L/min of O<sub>2</sub> flow, the mean ± SD FiO<sub>2</sub> was 29 ± 6; 39 ± 9 and 48 ± 2% with NC, NC_Sfm and NC_Dtm, respectively (p<0.01). From 5 to 9 L/min of O2, the mean ± SD FiO<sub>2</sub> was 45 ± 11; 59 ± 13; 69 ± 13 and 41 ± 8% with NC, NC_Sfm and NC_Dtm and FM, respectively (p<0.01). At 10 L/min, the NC_Dtm significantly outperformed all other systems (FiO<sub>2</sub> = 69 ± 13%) and was followed by the NRM (59 ± 9%). The results remained consistent regardless of the mouth condition or the set breathing pattern. <bold>Conclusions:</bold> For a given oxygen flow rate, FiO<sub>2</sub> values delivered by NC_Dtm was higher than those obtained with other oxygen delivery systems. This statement remains true regardless of the condition studied.
Authors
- Emeline Fresnel (ORCID: https://orcid.org/0000-0002-3011-7694)
- Roberto Martínez-Alejos (ORCID: https://orcid.org/0000-0002-2444-0575)
- William Poncin (ORCID: https://orcid.org/0000-0002-7686-2294)
- Shani Arrouch
Institutions
- Centre Hospitalier Universitaire de Rouen (FR)
- Université de Rouen Normandie (FR)
- UCLouvain (BE)
Publication Details
- Published
- 2024-09-14
- DOI
- https://doi.org/10.1183/13993003.congress-2024.pa1686
- Primary Topic
- Respiratory Support and Mechanisms
- Type
- article
- Field-Weighted Citation Impact
- 0.00