Does Use Of A Heat and Moisture Exchanger Affect pMDI Delivery to a Simulated Patient on Mechanical Ventilation

<bold>Objective:</bold> Heat and moisture exchangers (HME) have been designed to allow aerosol delivery to mechanically ventilated patients by bypassing the HME during aerosol administration. This study evaluates the effect a bypass-type HME has on aerosol delivery in a simulated adult ventilator setting. <bold>Methods:</bold> A Fisher & Paykel RT210 adult breathing circuit was used to simulate an adult model (500mL, duty cycle = 33%, 13 bpm) generated using a Draeger Infinity C500 ventilator. A Gibeck Humid-Flo HME was placed at the entrance to an 8.0mm diameter endotracheal tube (ETT). An aerosol collection filter was located at the distal end of the ETT and coupled to a SelfTestLung, simulating the patient. 5 actuations of Ventolin pMDI were delivered through 3 different delivery devices and a built-in port adapter in the circuit, followed by 6 complete breathing cycles. Salbutamol assay was undertaken by HPLC-UV. Comparisons were made on potential dose to the lungs and were equated to a potential relative carbon footprint based upon published claims [1] for Ventolin. <bold>Results:</bold> <bold>Conclusions:</bold> This study has shown that use of a bypass-type HME can allow delivery of medication to the patient. It also demonstrates that use of AeroChamber* VENT HC spacer could potentially reduce the carbon footprint by up to 4-fold compared to the alternative options. By maximizing the amount of each puff reaching the lungs the patient is likely to get relief sooner and reduce the number of puffs needed. <table-wrap><object-id>erj;64/suppl_68/PA2613/TB1</object-id><object-id>T1</object-id><object-id>TB1</object-id><table><colgroup><col></col><col></col></colgroup><tbody><tr><td>Device</td><td>Total Mass of Salbutamol /Actuation (mean ± SD)</td></tr><tr><td>AeroChamber* VENT HC</td><td>33.1±2.7µg</td></tr><tr><td>Spirale DDS</td><td>7.7±3.0µg</td></tr><tr><td>Hudson RCI MDI Adaptor</td><td>20.2±3.9µg</td></tr><tr><td>Built-in circuit pMDI port</td><td>22.9±1.2µg</td></tr></tbody></table></table-wrap> [1] https://greeninhaler.org

Authors

Institutions

Publication Details

Published
2024-09-14
DOI
https://doi.org/10.1183/13993003.congress-2024.pa2613
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Does Use Of A Heat and Moisture Exchanger Affect pMDI Delivery to a Simulated Patient on Mechanical Ventilation

Mark Nagel, Cathy Doyle, Jason Suggett, Rubina Ali
Respiratory Support and Mechanisms
article

Does Use Of A Heat and Moisture Exchanger Affect pMDI Delivery to a Simulated Patient on Mechanical Ventilation

Mark Nagel, Cathy Doyle, Jason Suggett, Rubina Ali
article en

Abstract

<bold>Objective:</bold> Heat and moisture exchangers (HME) have been designed to allow aerosol delivery to mechanically ventilated patients by bypassing the HME during aerosol administration. This study evaluates the effect a bypass-type HME has on aerosol delivery in a simulated adult ventilator setting. <bold>Methods:</bold> A Fisher & Paykel RT210 adult breathing circuit was used to simulate an adult model (500mL, duty cycle = 33%, 13 bpm) generated using a Draeger Infinity C500 ventilator. A Gibeck Humid-Flo HME was placed at the entrance to an 8.0mm diameter endotracheal tube (ETT). An aerosol collection filter was located at the distal end of the ETT and coupled to a SelfTestLung, simulating the patient. 5 actuations of Ventolin pMDI were delivered through 3 different delivery devices and a built-in port adapter in the circuit, followed by 6 complete breathing cycles. Salbutamol assay was undertaken by HPLC-UV. Comparisons were made on potential dose to the lungs and were equated to a potential relative carbon footprint based upon published claims [1] for Ventolin. <bold>Results:</bold> <bold>Conclusions:</bold> This study has shown that use of a bypass-type HME can allow delivery of medication to the patient. It also demonstrates that use of AeroChamber* VENT HC spacer could potentially reduce the carbon footprint by up to 4-fold compared to the alternative options. By maximizing the amount of each puff reaching the lungs the patient is likely to get relief sooner and reduce the number of puffs needed. <table-wrap><object-id>erj;64/suppl_68/PA2613/TB1</object-id><object-id>T1</object-id><object-id>TB1</object-id><table><colgroup><col></col><col></col></colgroup><tbody><tr><td>Device</td><td>Total Mass of Salbutamol /Actuation (mean ± SD)</td></tr><tr><td>AeroChamber* VENT HC</td><td>33.1±2.7µg</td></tr><tr><td>Spirale DDS</td><td>7.7±3.0µg</td></tr><tr><td>Hudson RCI MDI Adaptor</td><td>20.2±3.9µg</td></tr><tr><td>Built-in circuit pMDI port</td><td>22.9±1.2µg</td></tr></tbody></table></table-wrap> [1] https://greeninhaler.org

Trudell Medical International (Canada) (CA)
Affordable and clean energy
Openalex Percentile: Top 63%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Does Use Of A Heat and Moisture Exchanger Affect pMDI Delivery to a Simulated Patient on Mechanical Ventilation — Mark Nagel, Cathy Doyle, et al. · (2024) | TGRS Research Map | TGRS