Accuracy of real-time data provided by mechanical insufflation-exsufflation devices

Mechanical in-exsufflation (MIE) aims to generate cough peak expiratory flows (PEF) similar to those of normal cough. Currently, these devices display PEF and inspiratory volume (Vti) in real time, but there is no data on whether these values correspond to the actual values generated. A bench study was conducted using a Michigan mechanical lung connected to 3 different MIE devices. We simulated normal and reduced compliance (60 and 30 ml/cmH2O). MIE settings were +20/-20, +30/-30, +40/-40, +40/-50, +40/-60, and +40/-70cmH2O. Values displayed on the screen were recorded cycle by cycle, while a pneumotachograph connected to the mechanical lung measured actual PEF and Vti. All devices underestimated PEF, with error being statistically higher for devices B and C (A -20.6 ± 4.9 L/min vs B -68.4 ± 10.8 L/min vs C -70.6 ± 18.9; p < 0.0001 both). Similarly, all devices underestimated Vti by an average of 410 mL. <fig><object-id>erj;64/suppl_68/PA1435/F1</object-id><object-id>F1</object-id><object-id>F1</object-id><graphic></graphic></fig> Multivariate analysis showed that reasons for PEF (adjr2=0.86; p<0.0001) and Vti (adjr2=0.70; p<0.0001) estimation error were type of device, high pressure levels and compliance. <fig><object-id>erj;64/suppl_68/PA1435/F2</object-id><object-id>F2</object-id><object-id>F2</object-id><graphic></graphic></fig> Real time monitoring displayed on MIE devices underestimates the PEF and Vti.

Authors

Institutions

Publication Details

Published
2024-09-14
DOI
https://doi.org/10.1183/13993003.congress-2024.pa1435
Primary Topic
Engineering Technology and Methodologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of real-time data provided by mechanical insufflation-exsufflation devices

Roberto Martínez Alejos, Marius Lebret, Alice Vuillermoz, Emeline Fresnel et al.
Engineering Technology and Methodologies
article

Accuracy of real-time data provided by mechanical insufflation-exsufflation devices

Roberto Martínez Alejos, Marius Lebret, Alice Vuillermoz, Emeline Fresnel, François Beloncle
article en

Abstract

Mechanical in-exsufflation (MIE) aims to generate cough peak expiratory flows (PEF) similar to those of normal cough. Currently, these devices display PEF and inspiratory volume (Vti) in real time, but there is no data on whether these values correspond to the actual values generated. A bench study was conducted using a Michigan mechanical lung connected to 3 different MIE devices. We simulated normal and reduced compliance (60 and 30 ml/cmH2O). MIE settings were +20/-20, +30/-30, +40/-40, +40/-50, +40/-60, and +40/-70cmH2O. Values displayed on the screen were recorded cycle by cycle, while a pneumotachograph connected to the mechanical lung measured actual PEF and Vti. All devices underestimated PEF, with error being statistically higher for devices B and C (A -20.6 ± 4.9 L/min vs B -68.4 ± 10.8 L/min vs C -70.6 ± 18.9; p < 0.0001 both). Similarly, all devices underestimated Vti by an average of 410 mL. <fig><object-id>erj;64/suppl_68/PA1435/F1</object-id><object-id>F1</object-id><object-id>F1</object-id><graphic></graphic></fig> Multivariate analysis showed that reasons for PEF (adjr2=0.86; p<0.0001) and Vti (adjr2=0.70; p<0.0001) estimation error were type of device, high pressure levels and compliance. <fig><object-id>erj;64/suppl_68/PA1435/F2</object-id><object-id>F2</object-id><object-id>F2</object-id><graphic></graphic></fig> Real time monitoring displayed on MIE devices underestimates the PEF and Vti.

Anton Proksch Institut (AT), Centre Hospitalier Universitaire de Rouen (FR), Université d'Angers (FR)
Affordable and clean energy
Openalex Percentile: Top 72%
Engineering Technology and Methodologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.