A Computational Study on Effects of PID Temperature Target and RF Frequency for PID‐Controlled Nonablative RF Cosmetic Systems
PID controller has ensured treatment safety and uniformity, in exchange for some efficiency. Among 12 parameter combinations, the one with a temperature of 70°C and RF frequency of 1.0 MHz achieved the highest thermal damage volume, which could potentially result in the best esthetic effect. Considering users' different susceptibility to heat, engineers or physicians can select better temperature targets and RF frequencies to bring the desired cosmetic results based on thermal damage volume curves from this study.