Analysis of the ion mobility spectra of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol

Objectives . To determine the ion mobilities of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol; the structure of ions corresponding to characteristic signals; the detection limits of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol with the Kerber-T ion drift detector and the Segment automatic stationary gas detector. Methods . Ion mobility spectrometry was used in order to determine the ion mobilities and detect analytes. The enthalpies of reactions of ion formation were calculated using the ORCA 4.1.1 software by means of the B3LYP density functional method with the 6-31G(d,p) basis set. Results . The ion mobilities of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol were determined. A method for recording ion mobility spectra and their mathematical processing was developed. The dependencies of the change in ion mobility spectra on the analyte concentration were also studied. Possible mechanisms were proposed for the formation of the ion mobility spectra observed, in accordance with the ionization features of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol. The enthalpies of ion formation were calculated. The ionization schemes of the compounds were shown. The generalized results of experimental studies were presented, as were the features of compound identification taking into account the structure of the spectra, the concentrations of substances, and the detection conditions. Conclusions . Characteristic signals of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol were identified. All studied hazardous substances can be detected with an ion mobility spectrometer at concentrations at the ppm level. The following detection limits of the substances were determined with the Segment gas detector: chloroacetophenone, 245 mg/m 3 ; tris(2-chloroethyl)amine, 0.01 mg/m 3 ; and methanethiol, 0.8 mg/m 3 .

Authors

Institutions

Publication Details

Journal
Fine Chemical Technologies
Published
2024-11-04
DOI
https://doi.org/10.32362/2410-6593-2024-19-5-462-478
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis of the ion mobility spectra of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol

Y R Shaltaeva, A. A. Fenin, Е. S. Osinova, Е. П. Баберкина et al.
Fine Chemical Technologies
Mass Spectrometry Techniques and Applications
article

Analysis of the ion mobility spectra of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol

Y R Shaltaeva, A. A. Fenin, Е. S. Osinova, Е. П. Баберкина, V.V. Belyakov, Т. B. Melamed, А. E. Kovalenko, D. А. Aleksandrova, G. V. Tsaplin, Roman Yakushin, Yuri B. Sinkevich, A. O. Shablya, A. G. Sazonov, L. A. Luzenina, A. A. Kaplin
article en

Abstract

Objectives . To determine the ion mobilities of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol; the structure of ions corresponding to characteristic signals; the detection limits of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol with the Kerber-T ion drift detector and the Segment automatic stationary gas detector. Methods . Ion mobility spectrometry was used in order to determine the ion mobilities and detect analytes. The enthalpies of reactions of ion formation were calculated using the ORCA 4.1.1 software by means of the B3LYP density functional method with the 6-31G(d,p) basis set. Results . The ion mobilities of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol were determined. A method for recording ion mobility spectra and their mathematical processing was developed. The dependencies of the change in ion mobility spectra on the analyte concentration were also studied. Possible mechanisms were proposed for the formation of the ion mobility spectra observed, in accordance with the ionization features of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol. The enthalpies of ion formation were calculated. The ionization schemes of the compounds were shown. The generalized results of experimental studies were presented, as were the features of compound identification taking into account the structure of the spectra, the concentrations of substances, and the detection conditions. Conclusions . Characteristic signals of chloroacetophenone, tris(2-chloroethyl)amine, and methanethiol were identified. All studied hazardous substances can be detected with an ion mobility spectrometer at concentrations at the ppm level. The following detection limits of the substances were determined with the Segment gas detector: chloroacetophenone, 245 mg/m 3 ; tris(2-chloroethyl)amine, 0.01 mg/m 3 ; and methanethiol, 0.8 mg/m 3 .

Fine Chemical TechnologiesVol. 19(5)
D. Mendeleyev University of Chemical Technology of Russia (RU), Modus Therapeutics (Sweden) (SE), National Research Nuclear University MEPhI (RU)
D. Mendeleev University of Chemical Technology of Russia
Clean water and sanitation
Openalex Percentile: Top 85%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.