In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types

In phantom in-stent restenosis in 8 mm stents, we observed similar full width at half-maximum for photon-counting detector CT and energy-integrating detector CT in 0% and 75% in-stent restenosis, but at 95% in-stent restenosis, FWHM tended to be more accurate in smaller slice thicknesses in both scanners. Subjective image assessment yielded best results at 0.2 mm slice thickness in photon-counting detector CT despite lower contrast-to-noise ratio.

Authors

Institutions

Publication Details

Journal
CardioVascular and Interventional Radiology
Published
2024-11-05
DOI
https://doi.org/10.1007/s00270-024-03874-y
Citations
2
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.37

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types

Catharina Klausenitz, Sven R. Hauck, Christian Loewe, Melanie Schernthaner et al.
2 citations
CardioVascular and Interventional Radiology
Advanced X-ray and CT Imaging
0.37
article

In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types

Catharina Klausenitz, Sven R. Hauck, Christian Loewe, Melanie Schernthaner, Maximilian Korbinian Michael Hoffner, Martin Funovics, Andreas Strassl, Theresa-Marie Dachs, Irene Steiner, Hossam M. Abdel-Rahman, Maximilian Kern, Albert Hannover
article en
2 citations

Abstract

In phantom in-stent restenosis in 8 mm stents, we observed similar full width at half-maximum for photon-counting detector CT and energy-integrating detector CT in 0% and 75% in-stent restenosis, but at 95% in-stent restenosis, FWHM tended to be more accurate in smaller slice thicknesses in both scanners. Subjective image assessment yielded best results at 0.2 mm slice thickness in photon-counting detector CT despite lower contrast-to-noise ratio.

CardioVascular and Interventional RadiologyVol. 48(1)
Medical University of Vienna (AT)
Universität Wien, Medizinische Universität Wien
Affordable and clean energy
Openalex Percentile: Top 42%
Advanced X-ray and CT Imaging
0.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types — Catharina Klausenitz, Sven R. Hauck, et al. · CardioVascular and Interventional Radiology (2024) | TGRS Research Map | TGRS