In-Stent Restenosis in Peripheral Arterial Disease: Ultra-High-Resolution Photon-Counting Versus Third-Generation Dual-Source Energy-Integrating Detector CT Phantom Study in Seven Different Stent Types
In phantom in-stent restenosis in 8 mm stents, we observed similar full width at half-maximum for photon-counting detector CT and energy-integrating detector CT in 0% and 75% in-stent restenosis, but at 95% in-stent restenosis, FWHM tended to be more accurate in smaller slice thicknesses in both scanners. Subjective image assessment yielded best results at 0.2 mm slice thickness in photon-counting detector CT despite lower contrast-to-noise ratio.