Analytical Characteristics and Clinical Performance of Anti-Müllerian Hormone Immunoassay on the ADVIA Centaur® System: A Comparison with Other Chemiluminescent Methods

Background: Serum anti-Müllerian hormone (AMH) levels and antral follicle count are key in evaluating ovarian reserve (OR) for fertility. The performance of the Siemens Healthineers AMH assay was assessed on the ADVIA Centaur® System. Methods: Analytical characteristics, clinical performance, and method comparison studies were performed in a prospective cohort of 532 women at fertility clinics. Serum AMH levels were determined using ADVIA Centaur, Beckman Access®, and Roche Elecsys® assays. Results: The limit of quantitation for the ADVIA Centaur AMH assay was 0.030 ng/mL. Repeatability was ≤2.9% CV, within-lab repeatability was ≤3.2% CV, and reproducibility was ≤4.4% CV. Results using serum or lithium heparin sample types were equivalent. Diagnostic sensitivity across assays ranged from 77.3% to 90.2% and specificity ranged from 51.0 to 71.0%; corresponding positive and negative predictive values ranged from 66.6% to 74.3% and 74.2% to 83.0%, respectively. Receiver operating characteristic analyses demonstrated that the assays have a high probability for discriminating between diminished–normal and high OR. ADVIA and Beckman assays agreed according to ADVIA = 1.00 × Beckman + 0.014 ng/mL, τ = 0.909, while a more modest correlation of ADVIA = 1.41 × Roche − 0.024 ng/mL, τ = 0.777 was observed with Roche assay. Conclusions: The ADVIA Centaur assay demonstrates acceptable analytical characteristics and clinical performance comparable to the Roche AMH assay and is essentially interchangeable with the Beckman AMH assay for reliable OR assessment.

Authors

Institutions

Publication Details

Journal
Endocrines
Published
2024-11-05
DOI
https://doi.org/10.3390/endocrines5040037
Primary Topic
Ovarian function and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytical Characteristics and Clinical Performance of Anti-Müllerian Hormone Immunoassay on the ADVIA Centaur® System: A Comparison with Other Chemiluminescent Methods

Edward A. Zbella, Robert H. Christenson, Robert T. Singleton, Millie A. Behera et al.
Endocrines
Ovarian function and disorders
article

Analytical Characteristics and Clinical Performance of Anti-Müllerian Hormone Immunoassay on the ADVIA Centaur® System: A Comparison with Other Chemiluminescent Methods

Edward A. Zbella, Robert H. Christenson, Robert T. Singleton, Millie A. Behera, Jelena Bogdanović Pristov, Karoline Freeman, Caroline Brown, Jeanne E. O’Brien
article en

Abstract

Background: Serum anti-Müllerian hormone (AMH) levels and antral follicle count are key in evaluating ovarian reserve (OR) for fertility. The performance of the Siemens Healthineers AMH assay was assessed on the ADVIA Centaur® System. Methods: Analytical characteristics, clinical performance, and method comparison studies were performed in a prospective cohort of 532 women at fertility clinics. Serum AMH levels were determined using ADVIA Centaur, Beckman Access®, and Roche Elecsys® assays. Results: The limit of quantitation for the ADVIA Centaur AMH assay was 0.030 ng/mL. Repeatability was ≤2.9% CV, within-lab repeatability was ≤3.2% CV, and reproducibility was ≤4.4% CV. Results using serum or lithium heparin sample types were equivalent. Diagnostic sensitivity across assays ranged from 77.3% to 90.2% and specificity ranged from 51.0 to 71.0%; corresponding positive and negative predictive values ranged from 66.6% to 74.3% and 74.2% to 83.0%, respectively. Receiver operating characteristic analyses demonstrated that the assays have a high probability for discriminating between diminished–normal and high OR. ADVIA and Beckman assays agreed according to ADVIA = 1.00 × Beckman + 0.014 ng/mL, τ = 0.909, while a more modest correlation of ADVIA = 1.41 × Roche − 0.024 ng/mL, τ = 0.777 was observed with Roche assay. Conclusions: The ADVIA Centaur assay demonstrates acceptable analytical characteristics and clinical performance comparable to the Roche AMH assay and is essentially interchangeable with the Beckman AMH assay for reliable OR assessment.

EndocrinesVol. 5(4)
University of Maryland, Baltimore (US), Siemens Healthcare (United States) (US), Reproductive Genetics Institute (US), Shady Grove Fertility Center (US)
Openalex Percentile: Top 70%
Ovarian function and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.