A case study on the effect of contaminated inlet tubes on the accuracy of mid-cost optical particle counters for the ambient air monitoring of fine particles

Abstract This study investigates the impact of the long-term use of inlet-heated tubes on the performance of mid-cost optical particle counters (OPCs) for ambient air monitoring of fine particles (PM 2.5 ). Two OPCs, equipped with inlet-heated tubes, were deployed over a 6-month period, with a beta attenuation monitor (BAM) serving as the reference device. The performance of the OPCs using the same inlet tubes for the first 3 months was compared to their performance after the frequent replacement of clean tubes during the final 3 months. The correlation coefficients ( r 2 ) for the 1 h and 24 h average PM 2.5 concentrations between the OPCs and the BAM were lower with long-term contaminated tubes (0.82 < r 2 < 0.93) compared to clean tubes ( r 2 > 0.93). The relative mean errors and biases significantly increased over time with contaminated tubes. Temperature, humidity, precipitation, and wind speed were found to have an insignificant effect ( r 2 < 0.1) on the performance of the two OPCs with inlet-heated tubes over the 6-month period. The relative average PM 2.5 error when using clean tubes was less than 4%. These findings highlight the importance of inlet-heated tubes in improving OPC performance, particularly for mitigating humidity effects. Graphical Abstract

Authors

Institutions

Publication Details

Journal
Asian Journal of Atmospheric Environment
Published
2024-11-05
DOI
https://doi.org/10.1007/s44273-024-00045-w
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A case study on the effect of contaminated inlet tubes on the accuracy of mid-cost optical particle counters for the ambient air monitoring of fine particles

Jo-Chun Kim, Trieu-Vuong Dinh, Da-Hyun Baek, Byeong-Gyu Park et al.
Asian Journal of Atmospheric Environment
Air Quality Monitoring and Forecasting
article

A case study on the effect of contaminated inlet tubes on the accuracy of mid-cost optical particle counters for the ambient air monitoring of fine particles

Jo-Chun Kim, Trieu-Vuong Dinh, Da-Hyun Baek, Byeong-Gyu Park, Sang‐Woo Lee, In‐Young Choi
article en

Abstract

Abstract This study investigates the impact of the long-term use of inlet-heated tubes on the performance of mid-cost optical particle counters (OPCs) for ambient air monitoring of fine particles (PM 2.5 ). Two OPCs, equipped with inlet-heated tubes, were deployed over a 6-month period, with a beta attenuation monitor (BAM) serving as the reference device. The performance of the OPCs using the same inlet tubes for the first 3 months was compared to their performance after the frequent replacement of clean tubes during the final 3 months. The correlation coefficients ( r 2 ) for the 1 h and 24 h average PM 2.5 concentrations between the OPCs and the BAM were lower with long-term contaminated tubes (0.82 < r 2 < 0.93) compared to clean tubes ( r 2 > 0.93). The relative mean errors and biases significantly increased over time with contaminated tubes. Temperature, humidity, precipitation, and wind speed were found to have an insignificant effect ( r 2 < 0.1) on the performance of the two OPCs with inlet-heated tubes over the 6-month period. The relative average PM 2.5 error when using clean tubes was less than 4%. These findings highlight the importance of inlet-heated tubes in improving OPC performance, particularly for mitigating humidity effects. Graphical Abstract

Asian Journal of Atmospheric EnvironmentVol. 18(1)
Konkuk University (KR)
Ministry of Environment, Ministry of Education, India, Korea Environmental Industry and Technology Institute
Affordable and clean energy
Openalex Percentile: Top 83%
Air Quality Monitoring and Forecasting
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A case study on the effect of contaminated inlet tubes on the accuracy of mid-cost optical particle counters for the ambient air monitoring of fine particles — Jo-Chun Kim, Trieu-Vuong Dinh, et al. · Asian Journal of Atmospheric Environment (2024) | TGRS Research Map | TGRS