Effect of Ag–Te alloys on pH sensitivity of Ag2O–TeO2 glass/stainless steel enamel

Glass pH sensors are unsuitable for in vivo biomedical, clinical, or food applications because of brittleness of glass, the difficulty in measurement of small volumes and complex matrices, and the need for calibration. Herein, a novel enamel reference electrode for pH sensors is developed using a binary x Ag 2 O·(100– x )TeO 2 glass/stainless steel (SUS) electrode. Subsequently, the effects of Ag 2 O content on the precipitated crystalline phases and pH sensitivity of Ag 2 O–TeO 2 glass/SUS electrodes are examined. The results indicate that the x Ag 2 O·(100– x )TeO 2 glass/SUS samples exhibit relatively low pH sensitivity. The pH sensitivities of the samples can be classified into two distinct groups. The first group, comprising samples with an Ag 2 O content of 25 mol% or more, exhibits a pH sensitivity higher than 10 %. The second group, comprising samples with an Ag 2 O content of 22 mol% or less, exhibits a pH sensitivity of less than 10 %. The pH sensitivity tends to decrease with increase in stutzite fraction in Ag–Te alloys. Results of X-ray diffraction measurements and transmission electron microscopy observations indicate that the hessite and stutzite Ag–Te alloys, which are minor and conductive, are the primary contributors to the low pH sensitivity observed. Moreover, the pH sensitivity of x Ag 2 O·(100– x )TeO 2 /SUS is observed to be between those of the Ag–Te alloy and the Ag 2 O–TeO 2 glass matrix. We propose occurrence of an electrical shunt should occur between the Ag–Te alloy and the Ag + -ion conductive Ag 2 O–TeO 2 glass matrix. Regarding x Ag 2 O·(100– x )TeO 2 glass/SUS, the 20Ag 2 O·80TeO 2 glass/SUS to 21Ag 2 O·79TeO 2 glass/SUS compositions represent promising candidates as for KCl-leakage-free reference electrodes for pH sensors.

Authors

Institutions

Publication Details

Journal
Results in Chemistry
Published
2024-11-05
DOI
https://doi.org/10.1016/j.rechem.2024.101889
Citations
1
Primary Topic
Pigment Synthesis and Properties
Type
article
Field-Weighted Citation Impact
0.16

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Ag–Te alloys on pH sensitivity of Ag2O–TeO2 glass/stainless steel enamel

Atsushi Ishihara, Tomonari Kuno, Tadanori Hashimoto, Takuhisa Komi et al.
1 citations
Results in Chemistry
Pigment Synthesis and Properties
0.16
article

Effect of Ag–Te alloys on pH sensitivity of Ag2O–TeO2 glass/stainless steel enamel

Atsushi Ishihara, Tomonari Kuno, Tadanori Hashimoto, Takuhisa Komi, Yuji Nishio
article en
1 citations

Abstract

Glass pH sensors are unsuitable for in vivo biomedical, clinical, or food applications because of brittleness of glass, the difficulty in measurement of small volumes and complex matrices, and the need for calibration. Herein, a novel enamel reference electrode for pH sensors is developed using a binary x Ag 2 O·(100– x )TeO 2 glass/stainless steel (SUS) electrode. Subsequently, the effects of Ag 2 O content on the precipitated crystalline phases and pH sensitivity of Ag 2 O–TeO 2 glass/SUS electrodes are examined. The results indicate that the x Ag 2 O·(100– x )TeO 2 glass/SUS samples exhibit relatively low pH sensitivity. The pH sensitivities of the samples can be classified into two distinct groups. The first group, comprising samples with an Ag 2 O content of 25 mol% or more, exhibits a pH sensitivity higher than 10 %. The second group, comprising samples with an Ag 2 O content of 22 mol% or less, exhibits a pH sensitivity of less than 10 %. The pH sensitivity tends to decrease with increase in stutzite fraction in Ag–Te alloys. Results of X-ray diffraction measurements and transmission electron microscopy observations indicate that the hessite and stutzite Ag–Te alloys, which are minor and conductive, are the primary contributors to the low pH sensitivity observed. Moreover, the pH sensitivity of x Ag 2 O·(100– x )TeO 2 /SUS is observed to be between those of the Ag–Te alloy and the Ag 2 O–TeO 2 glass matrix. We propose occurrence of an electrical shunt should occur between the Ag–Te alloy and the Ag + -ion conductive Ag 2 O–TeO 2 glass matrix. Regarding x Ag 2 O·(100– x )TeO 2 glass/SUS, the 20Ag 2 O·80TeO 2 glass/SUS to 21Ag 2 O·79TeO 2 glass/SUS compositions represent promising candidates as for KCl-leakage-free reference electrodes for pH sensors.

Results in ChemistryVol. 12
Horiba (Japan) (JP), Mie University (JP)
Japan Society for the Promotion of Science London
Openalex Percentile: Top 56%
Pigment Synthesis and Properties
0.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.