A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds

Coronary artery disease (CAD) is characterized by narrowing and subsequent blockade of coronary arteries, and imposes a significant health and economic burden. Stent and scaffold devices are introduced in advanced CAD to improve vascular stability and restore blood flow. Although in vitro flow systems like the Chandler loop have been developed to enhance the understanding of interactions between device materials, their coatings, and vascular cells, imaging-based in vitro analysis of device performance is limited. In this study, we established a novel stent flow chamber system designed to assess the thrombogenicity of bioresorbable magnesium scaffold (RMS) and stent materials in vitro. Additionally, we compared the thrombogenicity - an important clinical parameter in stent performance - of the Magmaris-316 L stainless steel stent with its predecessors, Magmaris RMS and a prototype of the third-generation RMS (DREAMS 3G). Analysis of platelet adhesion and coverage of the different devices under flow conditions demonstrated that the Magmaris RMS exhibits reduced thrombogenicity compared to the Magmaris-316 L stainless steel stent. Moreover, thrombogenicity of the DREAMS 3G prototype, composed of BIOmag material, is further decreased compared to its predecessors. The observed reduction in thrombogenicity of the DREAMS 3G prototype in vitro suggests additional improvements in clinical safety and efficacy, highlighting its promise for treating CAD. Future research on this prototype may thus open avenues for analyzing other blood components and patient-derived endothelial cells. In line with the 3R principles, this approach may also help reduce the need for animal testing.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-11-04
DOI
https://doi.org/10.1038/s41598-024-77266-0
Citations
2
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
1.15

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds

S Lang, Maike Frye, Sandra A. Hemkemeyer, Manu Beerens et al.
2 citations
Scientific Reports
Coronary Interventions and Diagnostics
1.15
article

A novel stent flow chamber system demonstrates reduced thrombogenicity of bioresorbable magnesium scaffolds

S Lang, Maike Frye, Sandra A. Hemkemeyer, Manu Beerens, Marguerite Müller, Thomas Renné, Katharina Alina Riedl, Hanna Englert, Reiner K. Mailer, MJ Müller, Louis Ludwig, Philine Baumann‐Zumstein
article en
2 citations

Abstract

Coronary artery disease (CAD) is characterized by narrowing and subsequent blockade of coronary arteries, and imposes a significant health and economic burden. Stent and scaffold devices are introduced in advanced CAD to improve vascular stability and restore blood flow. Although in vitro flow systems like the Chandler loop have been developed to enhance the understanding of interactions between device materials, their coatings, and vascular cells, imaging-based in vitro analysis of device performance is limited. In this study, we established a novel stent flow chamber system designed to assess the thrombogenicity of bioresorbable magnesium scaffold (RMS) and stent materials in vitro. Additionally, we compared the thrombogenicity - an important clinical parameter in stent performance - of the Magmaris-316 L stainless steel stent with its predecessors, Magmaris RMS and a prototype of the third-generation RMS (DREAMS 3G). Analysis of platelet adhesion and coverage of the different devices under flow conditions demonstrated that the Magmaris RMS exhibits reduced thrombogenicity compared to the Magmaris-316 L stainless steel stent. Moreover, thrombogenicity of the DREAMS 3G prototype, composed of BIOmag material, is further decreased compared to its predecessors. The observed reduction in thrombogenicity of the DREAMS 3G prototype in vitro suggests additional improvements in clinical safety and efficacy, highlighting its promise for treating CAD. Future research on this prototype may thus open avenues for analyzing other blood components and patient-derived endothelial cells. In line with the 3R principles, this approach may also help reduce the need for animal testing.

Scientific ReportsVol. 14(1)
Royal College of Surgeons in Ireland (IE), Universität Hamburg (DE), Johannes Gutenberg University Mainz (DE), University Medical Center of the Johannes Gutenberg University Mainz (DE), University Medical Center Hamburg-Eppendorf (DE), German Centre for Cardiovascular Research (DE), Biotronik (Switzerland) (CH)
Deutsche Forschungsgemeinschaft, Universitätsklinikum Hamburg-Eppendorf
Good health and well-being
Openalex Percentile: Top 20%
Coronary Interventions and Diagnostics
1.15
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.