Measurements Of Arterial Occlusion Pressure Using Hand-held Devices

Use of blood flow restriction (BFR) during training has become increasingly popular due to the benefits over a wide range of applications. An essential component to the safe and effective use of BFR is the measurement of arterial occlusion pressure (AOP). PURPOSE: This study compared measures of AOP of the brachial artery using three devices and two cuff inflation methods. METHODS: Brachial artery AOP was measured in 20 males and 21 females simultaneously using Doppler ultrasound (US), a hand-held Doppler (HHDOP) and a pulse oximeter (PO) once when inflating the cuff with a clinical grade Hokanson (HOK) rapid cuff inflation system and twice manually (MAN) with a sphygmomanometer. RESULTS: A mixed model ANOVA revealed small but significant (p < 0.05) overall main effects between AOP measured using the HOK (120.4 ± 1.98) and MAN (122.2 ± 2.0) cuff inflation methods, between US(122.0 ± 1.97), HHDOP (121.6 ± 2.0) and PO (120.5 ± 2.0) measurements of AOP, and between males(127.6 ± 2.83) and females (115.2 ± 2.7). Further analyses indicated that the small overall difference between US and PO (1.56 ± 0.52) measures of AOP was significant (p < 0.05) but the overall differences between US and HHDOP (0.51 ± 0.52) measures of AOP was not significant (p > 0.05). Trial-to-trial variance in measures of AOP using US, HHDOP and PO were negligible. Bland-Altman plots revealed reasonable limits of agreement for both HHDOP (±4.46 mmHg) and PO (±5.47 mmHg) measures of AOP. CONCLUSIONS: The small differences in US, HHDOP and PO measures of AOP measurements using HOK and MAN cuff inflation methods are of little practical significance. Manual inflation of the pressure cuff provides comparable AOP values compared to when using a clinical grade cuff inflation system. Practitioners can be confident in measures of AOP using a quality hand-held doppler or pulse oximeter prior to blood flow restriction training.

Authors

Institutions

Publication Details

Journal
Medicine & Science in Sports & Exercise
Published
2024-10-01
DOI
https://doi.org/10.1249/01.mss.0001093636.69334.c1
Primary Topic
Peripheral Artery Disease Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurements Of Arterial Occlusion Pressure Using Hand-held Devices

Ryan Lacey, Pat R. Vehrs, Shay Richards, Chase Blazzard et al.
Medicine & Science in Sports & Exercise
Peripheral Artery Disease Management
article

Measurements Of Arterial Occlusion Pressure Using Hand-held Devices

Ryan Lacey, Pat R. Vehrs, Shay Richards, Chase Blazzard, Josh Allen, Hannah Hart, Tyler Burbank, Rachel Barrett, Daniela López, Josh Richard Nielsen
article en

Abstract

Use of blood flow restriction (BFR) during training has become increasingly popular due to the benefits over a wide range of applications. An essential component to the safe and effective use of BFR is the measurement of arterial occlusion pressure (AOP). PURPOSE: This study compared measures of AOP of the brachial artery using three devices and two cuff inflation methods. METHODS: Brachial artery AOP was measured in 20 males and 21 females simultaneously using Doppler ultrasound (US), a hand-held Doppler (HHDOP) and a pulse oximeter (PO) once when inflating the cuff with a clinical grade Hokanson (HOK) rapid cuff inflation system and twice manually (MAN) with a sphygmomanometer. RESULTS: A mixed model ANOVA revealed small but significant (p < 0.05) overall main effects between AOP measured using the HOK (120.4 ± 1.98) and MAN (122.2 ± 2.0) cuff inflation methods, between US(122.0 ± 1.97), HHDOP (121.6 ± 2.0) and PO (120.5 ± 2.0) measurements of AOP, and between males(127.6 ± 2.83) and females (115.2 ± 2.7). Further analyses indicated that the small overall difference between US and PO (1.56 ± 0.52) measures of AOP was significant (p < 0.05) but the overall differences between US and HHDOP (0.51 ± 0.52) measures of AOP was not significant (p > 0.05). Trial-to-trial variance in measures of AOP using US, HHDOP and PO were negligible. Bland-Altman plots revealed reasonable limits of agreement for both HHDOP (±4.46 mmHg) and PO (±5.47 mmHg) measures of AOP. CONCLUSIONS: The small differences in US, HHDOP and PO measures of AOP measurements using HOK and MAN cuff inflation methods are of little practical significance. Manual inflation of the pressure cuff provides comparable AOP values compared to when using a clinical grade cuff inflation system. Practitioners can be confident in measures of AOP using a quality hand-held doppler or pulse oximeter prior to blood flow restriction training.

Medicine & Science in Sports & ExerciseVol. 56(10S)
Brigham Young University (US)
Openalex Percentile: Top 63%
Peripheral Artery Disease Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.