Investigation of stent retriever removal forces in an experimental model of acute ischemic stroke
Benchtop mechanical thrombectomies performed with commercial stent retrievers of varying geometry showed that device removal forces increase with increasing model tortuosity, clot hematocrit does not play a significant role in device removal force, and that a stent retriever's diameter has a greater impact on removal forces compared to its length. These results provide an improved understanding of the overall forces involved in mechanical thrombectomy and can be used to develop safer and more effective stent retrievers for the most difficult cases.
Authors
- Bryan C. Good (ORCID: https://orcid.org/0000-0002-2662-8918)
- Michael T. Froehler (ORCID: https://orcid.org/0000-0003-1282-005X)
- Demitria A. Poulos
Institutions
- University of Tennessee at Knoxville (US)
- Vanderbilt University Medical Center (US)
Publication Details
- Journal
- Frontiers in Neurology
- Published
- 2024-10-31
- DOI
- https://doi.org/10.3389/fneur.2024.1486738
- Citations
- 5
- Primary Topic
- Acute Ischemic Stroke Management
- Type
- article
- Field-Weighted Citation Impact
- 1.93