Correlation between subretinal tissue plasminogen activator and air injection rates with pressure in a retina mimicking model

By investigating the correlation between the injection rate and pressure of subretinal tissue plasminogen activator (tPA) and air using a standard Viscous Fluid Control (VFC) system with a 38-gauge cannula, we aimed to establish guidelines for stable injections. We fabricated a retina mimicking model (RMM) with 0.25% agarose solution and an aluminum plate, and substituted submacular hemorrhage (SMH) and tPA with blood-mimicking fluid (BMF) and balanced salt solution (BSS), respectively. The diameter of the pre-bleb mimicking SMH in RMM was 1.30 ± 0.16 cm, increasing to 1.98 ± 0.24 cm and 1.83 ± 0.22 cm after bleb propagation with BSS and air, respectively. BSS injection rates were 2.86 ± 0.04 µl/sec, 6.74 ± 0.48 µl/sec and 8.55 ± 0.16 µl/sec at 8, 12, and 16 psi, respectively. Air injection rates were 37.98 ± 3.11 µl/sec, 79.01 ± 5.13 µl/sec and 156.06 ± 13.72 µl/sec at 2, 3 and 4 psi, respectively. By experimenting with different pressures in the RMM, we found 12 psi to be the minimum for proper BSS injection and 2 psi for air. These findings provide crucial parameters for safer surgery to prevent irreversible damage.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-10-31
DOI
https://doi.org/10.1038/s41598-024-77518-z
Primary Topic
Retinal Development and Disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correlation between subretinal tissue plasminogen activator and air injection rates with pressure in a retina mimicking model

Na Hee Kim, Minsub Lee, Hyewon Chung, Hyungwoo Lee et al.
Scientific Reports
Retinal Development and Disorders
article

Correlation between subretinal tissue plasminogen activator and air injection rates with pressure in a retina mimicking model

Na Hee Kim, Minsub Lee, Hyewon Chung, Hyungwoo Lee, Hyung Chan Kim
article en

Abstract

By investigating the correlation between the injection rate and pressure of subretinal tissue plasminogen activator (tPA) and air using a standard Viscous Fluid Control (VFC) system with a 38-gauge cannula, we aimed to establish guidelines for stable injections. We fabricated a retina mimicking model (RMM) with 0.25% agarose solution and an aluminum plate, and substituted submacular hemorrhage (SMH) and tPA with blood-mimicking fluid (BMF) and balanced salt solution (BSS), respectively. The diameter of the pre-bleb mimicking SMH in RMM was 1.30 ± 0.16 cm, increasing to 1.98 ± 0.24 cm and 1.83 ± 0.22 cm after bleb propagation with BSS and air, respectively. BSS injection rates were 2.86 ± 0.04 µl/sec, 6.74 ± 0.48 µl/sec and 8.55 ± 0.16 µl/sec at 8, 12, and 16 psi, respectively. Air injection rates were 37.98 ± 3.11 µl/sec, 79.01 ± 5.13 µl/sec and 156.06 ± 13.72 µl/sec at 2, 3 and 4 psi, respectively. By experimenting with different pressures in the RMM, we found 12 psi to be the minimum for proper BSS injection and 2 psi for air. These findings provide crucial parameters for safer surgery to prevent irreversible damage.

Scientific ReportsVol. 14(1)
Konkuk University Medical Center (KR)
Konkuk University
Openalex Percentile: Top 83%
Retinal Development and Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.