Intersystem and Interoperator Agreement of US Attenuation Coefficient for Quantifying Liver Steatosis

Background The extent of liver steatosis can be assessed using US attenuation coefficient (AC) algorithms currently implemented in several US systems. However, little is known about intersystem and interoperator variability in measurements. Purpose To assess intersystem and interoperator agreement in US AC measurements for fat quantification in individuals with varying degrees of liver steatosis and to assess the correlation of each manufacturer's AC algorithm results with MRI proton density fat fraction (PDFF). Materials and Methods This prospective study was conducted at Southwoods Imaging, Youngstown, Ohio, September 30-October 1, 2023. Two operators independently obtained AC measurements using eight US systems equipped with an AC algorithm from different manufacturers. On the same day, MRI PDFF measurement was performed by a different operator. Correlation between US AC and MRI PDFF was assessed using a mixed-effects model. Agreement between systems and operators was evaluated using the intraclass correlation coefficient (ICC). Results Twenty-six individuals (mean age, 55.4 years ± 10.7 [SD]; 16 female participants) were evaluated. The correlation of US AC with MRI PDFF was high for five AC algorithms (<i>r</i> range, 0.70-0.86), moderate for two (<i>r</i> = 0.62 for both), and poor for one (<i>r</i> = 0.47). In pairwise comparisons, none of the pairs of systems achieved excellent agreement (overall ICC = 0.33 [95% CI: 0.15, 0.52]). One pair showed good agreement (ICC = 0.79 [95% CI: 0.66, 0.87]), eight pairs showed moderate agreement (ICC range, 0.50 [95% CI: 0.22, 0.69] to 0.73 [95% CI: 0.49, 0.85]), and 19 pairs showed poor agreement (ICC range, 0.11 [95% CI: -0.06, 0.37] to 0.48 [95% CI: 0.20, 0.67]). Interoperator agreement on AC value was excellent for the Samsung Medison algorithm (ICC = 0.90 [95% CI: 0.80, 0.96]), good for the Siemens Healthineers (ICC = 0.76 [95% CI: 0.54, 0.89]) and Canon Medical Systems (ICC = 0.76 [95% CI: 0.16, 0.92]) algorithms, and moderate for the remaining algorithms (ICC range, 0.50 [95% CI: 0.16, 0.73] to 0.74 [95% CI: 0.51, 0.88]). The mean AC value obtained by the two operators did not differ for any system except the system from Canon Medical Systems. Conclusion There was substantial variability in AC values obtained with different US systems, precluding interchangeability between systems for liver steatosis diagnosis and follow-up imaging. Interoperator agreement ranged from moderate to excellent. © RSNA, 2024 <i>Supplemental material is available for this article.</i> See also the editorial by Han in this issue.

Authors

Institutions

Publication Details

Journal
Radiology
Published
2024-10-01
DOI
https://doi.org/10.1148/radiol.240162
Citations
12
Primary Topic
Liver Disease Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
4.64
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intersystem and Interoperator Agreement of US Attenuation Coefficient for Quantifying Liver Steatosis

Davide Roccarina, R. Graham Barr, Giovanna Ferraioli
12 citations
Radiology
Liver Disease Diagnosis and Treatment
4.64
article

Intersystem and Interoperator Agreement of US Attenuation Coefficient for Quantifying Liver Steatosis

Davide Roccarina, R. Graham Barr, Giovanna Ferraioli
article en
12 citations

Abstract

Background The extent of liver steatosis can be assessed using US attenuation coefficient (AC) algorithms currently implemented in several US systems. However, little is known about intersystem and interoperator variability in measurements. Purpose To assess intersystem and interoperator agreement in US AC measurements for fat quantification in individuals with varying degrees of liver steatosis and to assess the correlation of each manufacturer's AC algorithm results with MRI proton density fat fraction (PDFF). Materials and Methods This prospective study was conducted at Southwoods Imaging, Youngstown, Ohio, September 30-October 1, 2023. Two operators independently obtained AC measurements using eight US systems equipped with an AC algorithm from different manufacturers. On the same day, MRI PDFF measurement was performed by a different operator. Correlation between US AC and MRI PDFF was assessed using a mixed-effects model. Agreement between systems and operators was evaluated using the intraclass correlation coefficient (ICC). Results Twenty-six individuals (mean age, 55.4 years ± 10.7 [SD]; 16 female participants) were evaluated. The correlation of US AC with MRI PDFF was high for five AC algorithms (<i>r</i> range, 0.70-0.86), moderate for two (<i>r</i> = 0.62 for both), and poor for one (<i>r</i> = 0.47). In pairwise comparisons, none of the pairs of systems achieved excellent agreement (overall ICC = 0.33 [95% CI: 0.15, 0.52]). One pair showed good agreement (ICC = 0.79 [95% CI: 0.66, 0.87]), eight pairs showed moderate agreement (ICC range, 0.50 [95% CI: 0.22, 0.69] to 0.73 [95% CI: 0.49, 0.85]), and 19 pairs showed poor agreement (ICC range, 0.11 [95% CI: -0.06, 0.37] to 0.48 [95% CI: 0.20, 0.67]). Interoperator agreement on AC value was excellent for the Samsung Medison algorithm (ICC = 0.90 [95% CI: 0.80, 0.96]), good for the Siemens Healthineers (ICC = 0.76 [95% CI: 0.54, 0.89]) and Canon Medical Systems (ICC = 0.76 [95% CI: 0.16, 0.92]) algorithms, and moderate for the remaining algorithms (ICC range, 0.50 [95% CI: 0.16, 0.73] to 0.74 [95% CI: 0.51, 0.88]). The mean AC value obtained by the two operators did not differ for any system except the system from Canon Medical Systems. Conclusion There was substantial variability in AC values obtained with different US systems, precluding interchangeability between systems for liver steatosis diagnosis and follow-up imaging. Interoperator agreement ranged from moderate to excellent. © RSNA, 2024 <i>Supplemental material is available for this article.</i> See also the editorial by Han in this issue.

RadiologyVol. 313(1)
Azienda Ospedaliero-Universitaria Careggi (IT), University College London (GB)
Partnerships for the goals
Openalex Percentile: Top 4%
Liver Disease Diagnosis and Treatment
4.64
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.