[Research on motion impedance cardiography de-noising method based on two-step spectral ensemble empirical mode decomposition and canonical correlation analysis].

Impedance cardiography (ICG) is essential in evaluating cardiac function in patients with cardiovascular diseases. Aiming at the problem that the measurement of ICG signal is easily disturbed by motion artifacts, this paper introduces a de-noising method based on two-step spectral ensemble empirical mode decomposition (EEMD) and canonical correlation analysis (CCA). Firstly, the first spectral EEMD-CCA was performed between ICG and motion signals, and electrocardiogram (ECG) and motion signals, respectively. The component with the strongest correlation coefficient was set to zero to suppress the main motion artifacts. Secondly, the obtained ECG and ICG signals were subjected to a second spectral EEMD-CCA for further denoising. Lastly, the ICG signal is reconstructed using these share components. The experiment was tested on 30 subjects, and the results showed that the quality of the ICG signal is greatly improved after using the proposed denoising method, which could support the subsequent diagnosis and analysis of cardiovascular diseases.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2024-10-25
DOI
https://doi.org/10.7507/1001-5515.202210059
Primary Topic
Traditional Chinese Medicine Studies
Type
article
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

[Research on motion impedance cardiography de-noising method based on two-step spectral ensemble empirical mode decomposition and canonical correlation analysis].

Yao Xie, Qilian Xie, Honglong Yu, Dong Yang
PubMed
Traditional Chinese Medicine Studies
article

[Research on motion impedance cardiography de-noising method based on two-step spectral ensemble empirical mode decomposition and canonical correlation analysis].

Yao Xie, Qilian Xie, Honglong Yu, Dong Yang
article en

Abstract

Impedance cardiography (ICG) is essential in evaluating cardiac function in patients with cardiovascular diseases. Aiming at the problem that the measurement of ICG signal is easily disturbed by motion artifacts, this paper introduces a de-noising method based on two-step spectral ensemble empirical mode decomposition (EEMD) and canonical correlation analysis (CCA). Firstly, the first spectral EEMD-CCA was performed between ICG and motion signals, and electrocardiogram (ECG) and motion signals, respectively. The component with the strongest correlation coefficient was set to zero to suppress the main motion artifacts. Secondly, the obtained ECG and ICG signals were subjected to a second spectral EEMD-CCA for further denoising. Lastly, the ICG signal is reconstructed using these share components. The experiment was tested on 30 subjects, and the results showed that the quality of the ICG signal is greatly improved after using the proposed denoising method, which could support the subsequent diagnosis and analysis of cardiovascular diseases.

PubMedVol. 41(5)
University of Science and Technology of China (CN), Anhui Medical University (CN), Tongling Nonferrous Metals Group Holding (China) (CN)
Traditional Chinese Medicine Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.