0.35-mm lead-equivalent aprons provide similar protection to interventionalists compared to 0.5-mm lead-equivalent aprons during fluoroscopically guided interventions

<h2>Abstract</h2><h3>Objective</h3> Protective garments are part of the routine radiation safety equipment used during fluoroscopically guided interventions (FGIs). New elements and lead equivalences have been introduced by manufacturers, which may influence protection for interventionalists from ionizing radiation. We sought to determine the performance of different lead-equivalent apron vests during FGIs in clinical and simulated scenarios. <h3>Methods</h3> Our primary endpoint was to evaluate the effectiveness of radiation dose attenuation of two different protective leaded apron vests, one with nominal 0.5-mm lead equivalence and a second with 0.35-mm lead equivalence, which were evaluated in clinical and simulated settings. In the clinical setting, optically stimulated luminescence nanoDot™ detectors were placed at the upper outer quadrant (UOQ) chest wall (CW) position, both over and under the apron vests on one vascular surgeon performing FGIs over a 21-month period. All interventions were performed in hybrid rooms with Allura Clarity (Phillips Healthcare) C-arms with state of-the-art software for vessel navigation, digital subtraction angiography, digital magnification, and collimation. Dosimeters were placed on either side of the primary operator's body, located on the side closest to the X-ray source. In the simulation, fluorography was performed on a 30-inch-thick acrylic scatter phantom at 68, 80, 100, and 120 kVp for an exposure of 2000 mGy reference air kerma. Experiments were performed on the 0.35-mm and 0.5-mm lead-equivalent aprons. Paired Wilcoxon, χ<sup>2</sup>, and analysis of variance tests were performed to identify statistical significance of radiation attenuation dose rates. <h3>Results</h3> Operator UOQ CW radiation dose was measured during 32 FGIs: 16 were performed with the 0.5-mm lead-equivalent apron and 16 with the 0.35-mm lead-equivalent apron. Median procedure reference air kerma was 167 mGy (interquartile range, 99-437 mGy) when the 0.5-mm apron vest was worn vs 250 mGy (interquartile range, 144-410 mGy) with the 0.35-mm vest. There was no significant difference in UOQ CW radiation dose attenuation between the two lead equivalencies: thick 89% vs thin 86%; <i>P</i> = .2. In the simulated scenario, radiation dose attenuation was similar for all measured kVp, with no significant differences for both apron vests (94% thick vs 95% thin; <i>P</i> = .49). <h3>Conclusions</h3> Heavier leaded aprons do not offer clinically significant increased protection over thinner lead. Due to the long-term musculoskeletal strain on interventionalists, it is safe to consider lightweight lead protection.

Authors

Institutions

Publication Details

Journal
JVS-Vascular Insights
Published
2024-10-28
DOI
https://doi.org/10.1016/j.jvsvi.2024.100154
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

0.35-mm lead-equivalent aprons provide similar protection to interventionalists compared to 0.5-mm lead-equivalent aprons during fluoroscopically guided interventions

Antonio Solano, Melissa L. Kirkwood, Mirza Shadman Baig, Alejandro Pizano et al.
1 citations
JVS-Vascular Insights
Radiation Dose and Imaging
0.40
article

0.35-mm lead-equivalent aprons provide similar protection to interventionalists compared to 0.5-mm lead-equivalent aprons during fluoroscopically guided interventions

Antonio Solano, Melissa L. Kirkwood, Mirza Shadman Baig, Alejandro Pizano, Bala Ramanan, Khalil Chamseddin, Carlos H. Timaran, Gerardo Gonzalez-Guardiola, Michael Cyrus Siah, Michael Shih, Jeffrey Guild, Vivek Prakash
article en
1 citations

Abstract

<h2>Abstract</h2><h3>Objective</h3> Protective garments are part of the routine radiation safety equipment used during fluoroscopically guided interventions (FGIs). New elements and lead equivalences have been introduced by manufacturers, which may influence protection for interventionalists from ionizing radiation. We sought to determine the performance of different lead-equivalent apron vests during FGIs in clinical and simulated scenarios. <h3>Methods</h3> Our primary endpoint was to evaluate the effectiveness of radiation dose attenuation of two different protective leaded apron vests, one with nominal 0.5-mm lead equivalence and a second with 0.35-mm lead equivalence, which were evaluated in clinical and simulated settings. In the clinical setting, optically stimulated luminescence nanoDot™ detectors were placed at the upper outer quadrant (UOQ) chest wall (CW) position, both over and under the apron vests on one vascular surgeon performing FGIs over a 21-month period. All interventions were performed in hybrid rooms with Allura Clarity (Phillips Healthcare) C-arms with state of-the-art software for vessel navigation, digital subtraction angiography, digital magnification, and collimation. Dosimeters were placed on either side of the primary operator's body, located on the side closest to the X-ray source. In the simulation, fluorography was performed on a 30-inch-thick acrylic scatter phantom at 68, 80, 100, and 120 kVp for an exposure of 2000 mGy reference air kerma. Experiments were performed on the 0.35-mm and 0.5-mm lead-equivalent aprons. Paired Wilcoxon, χ<sup>2</sup>, and analysis of variance tests were performed to identify statistical significance of radiation attenuation dose rates. <h3>Results</h3> Operator UOQ CW radiation dose was measured during 32 FGIs: 16 were performed with the 0.5-mm lead-equivalent apron and 16 with the 0.35-mm lead-equivalent apron. Median procedure reference air kerma was 167 mGy (interquartile range, 99-437 mGy) when the 0.5-mm apron vest was worn vs 250 mGy (interquartile range, 144-410 mGy) with the 0.35-mm vest. There was no significant difference in UOQ CW radiation dose attenuation between the two lead equivalencies: thick 89% vs thin 86%; <i>P</i> = .2. In the simulated scenario, radiation dose attenuation was similar for all measured kVp, with no significant differences for both apron vests (94% thick vs 95% thin; <i>P</i> = .49). <h3>Conclusions</h3> Heavier leaded aprons do not offer clinically significant increased protection over thinner lead. Due to the long-term musculoskeletal strain on interventionalists, it is safe to consider lightweight lead protection.

JVS-Vascular InsightsVol. 3
Nassau University Medical Center (US), The University of Texas Southwestern Medical Center (US)
Zero hunger
Openalex Percentile: Top 33%
Radiation Dose and Imaging
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.