Analytic Performance of MicroVue Enzyme Immunoassay (EIA) for Measurement of Ba and sC5b-9 for Urine

Background: Complement mediated kidney diseases, including C3 glomerulopathy (C3G), atypical hemolytic uremic syndrome (aHUS), lupus nephritis (LN), and immunoglobulin A nephropathy (IgAN) are chronic inflammatory conditions characterized by dysregulated complement activation. Traditionally, complement activation is measured in plasma. However, patients with kidney disease often have an abundance of complement activation fragments in their urine – including Ba and soluble C5b-9 (sC5b-9), markers of alternative pathway activation and terminal pathway activation respectively– highlighting the potential value of Ba and sC5b-9 measurement in urine for diagnosis and risk management in patients with C3G, aHUS, LN, IgAN, and other kidney diseases. Herein we evaluated the analytical performance of the Quidel MicroVue Plus EIA Ba and sC5b-9 for urine samples. Methods: The Quidel MicroVue Ba and sC5b-9 Plus EIA were evaluated using randomly collected healthy urine samples or healthy urine samples spiked with a known concentrations of Ba and sC5b-9, as well as chronic kidney disease (CKD) urine samples. Results: The analytical performance of the assays met acceptability criteria for accuracy and precision. Assay accuracy was (100 ± 20%) and intra-assay precision as well as inter-assay precision was ≤ 20 % coefficient of variability (CV) for both assays. Ba was detected in healthy urine samples (N=40), but sC5b9 was not. Positive linearity was observed in both assays over a wide range of concentrations. Ba and sC5b-9 measured in urine was stable following three freeze-thaw cycles. Lastly, Ba and sC5b-9 measured in urine was stable when stored at -20°C and -80°C for 4 weeks. Long term stability testing at -80°C is underway. Conclusion: Quidel MicroVue Ba and sC5b-9 Plus EIA are accurate, and reliable assays for detection of Ba and sC5b-9 in urine samples, making it a viable alternatibe to plasma.

Authors

Institutions

Publication Details

Journal
Journal of the American Society of Nephrology
Published
2024-10-01
DOI
https://doi.org/10.1681/asn.2024041w8fyx
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analytic Performance of MicroVue Enzyme Immunoassay (EIA) for Measurement of Ba and sC5b-9 for Urine

Manuel Galvan, Eric Gorzkowski, Cameron Droste
Journal of the American Society of Nephrology
Biosensors and Analytical Detection
article

Analytic Performance of MicroVue Enzyme Immunoassay (EIA) for Measurement of Ba and sC5b-9 for Urine

Manuel Galvan, Eric Gorzkowski, Cameron Droste
article en

Abstract

Background: Complement mediated kidney diseases, including C3 glomerulopathy (C3G), atypical hemolytic uremic syndrome (aHUS), lupus nephritis (LN), and immunoglobulin A nephropathy (IgAN) are chronic inflammatory conditions characterized by dysregulated complement activation. Traditionally, complement activation is measured in plasma. However, patients with kidney disease often have an abundance of complement activation fragments in their urine – including Ba and soluble C5b-9 (sC5b-9), markers of alternative pathway activation and terminal pathway activation respectively– highlighting the potential value of Ba and sC5b-9 measurement in urine for diagnosis and risk management in patients with C3G, aHUS, LN, IgAN, and other kidney diseases. Herein we evaluated the analytical performance of the Quidel MicroVue Plus EIA Ba and sC5b-9 for urine samples. Methods: The Quidel MicroVue Ba and sC5b-9 Plus EIA were evaluated using randomly collected healthy urine samples or healthy urine samples spiked with a known concentrations of Ba and sC5b-9, as well as chronic kidney disease (CKD) urine samples. Results: The analytical performance of the assays met acceptability criteria for accuracy and precision. Assay accuracy was (100 ± 20%) and intra-assay precision as well as inter-assay precision was ≤ 20 % coefficient of variability (CV) for both assays. Ba was detected in healthy urine samples (N=40), but sC5b9 was not. Positive linearity was observed in both assays over a wide range of concentrations. Ba and sC5b-9 measured in urine was stable following three freeze-thaw cycles. Lastly, Ba and sC5b-9 measured in urine was stable when stored at -20°C and -80°C for 4 weeks. Long term stability testing at -80°C is underway. Conclusion: Quidel MicroVue Ba and sC5b-9 Plus EIA are accurate, and reliable assays for detection of Ba and sC5b-9 in urine samples, making it a viable alternatibe to plasma.

Journal of the American Society of NephrologyVol. 35(10S)
National Jewish Health (US), University of Colorado Denver (US)
Zero hunger
Openalex Percentile: Top 82%
Biosensors and Analytical Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Analytic Performance of MicroVue Enzyme Immunoassay (EIA) for Measurement of Ba and sC5b-9 for Urine — Manuel Galvan, Eric Gorzkowski, et al. · Journal of the American Society of Nephrology (2024) | TGRS Research Map | TGRS