Comparison of continuity index for drag versus point-by-point ablation with the use of proactive esophageal cooling

Abstract Background Proactive esophageal cooling was recently FDA cleared to reduce the likelihood of esophageal injury during radiofrequency (RF) ablation for treatment of atrial fibrillation (AF). Long-term follow-up data have also shown improved freedom from arrhythmia with proactive esophageal cooling compared to luminal esophageal temperature (LET) monitoring during pulmonary vein isolation (PVI). One hypothesized mechanism is the greater lesion continuity attainable with the use of cooling, which can be quantified by the Continuity Index (which increases as continuity decreases). Increasing the Continuity Index increases gap formation, reduces isolation, and leads to greater AF recurrence; however, the difference in Continuity Index between ablating in a point-by-point fashion versus a dragging fashion has not previously been analyzed. Purpose To prospectively measure the Continuity Index of PVI cases and compare the dragging technique to point-by-point technique when using proactive esophageal cooling or LET monitoring. Methods We determined the Continuity Index for 101 patients: 77 prospectively utilizing proactive esophageal cooling, 24 retrospectively recorded using LET monitoring. Cases were completed by four operators, three utilizing point-by-point technique, and one using the drag technique. The Continuity Index was calculated by incrementing one unit for each lesion that did not border the previous lesion during the PVI (simplifying the original description of the Continuity Index). Results For cases using esophageal cooling and performing RF ablation using the drag technique, the average Continuity Index was 8.5 (range: 1 to 14). For point-by-point operators using esophageal cooling, the average Continuity Index was 1.2 (range: 0 to 9), representing a difference of 7.3 between the dragging technique and point-by-point ablation (p < .001). Patients receiving LET monitoring and RF ablation using the drag technique had Continuity Indices of 23.8 (range 15 to 30). Patients receiving LET monitoring and RF ablation using point-by-point ablation had Continuity Indices of 28.4 (range 16 to 50). Figure 1 compares these findings. Only 4 patients (5%) in the esophageal cooling group had a Continuity Index above 6 on either side (all in the dragging group), versus 100% (24) of patients in the LET monitoring group. Conclusion Proactive esophageal cooling during PVI is associated with a high degree of lesion continuity, as represented by the low Continuity Index. The drag technique resulted in a higher index than the point-by-point technique, but both provided significantly lower Continuity Indices (greater lesion continuity) than cases using LET monitoring. This may offer a mechanism for the greater freedom from arrhythmia seen to date with proactive cooling in long-term follow-up.Figure 1.Comparison of Continuity Index

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.564
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of continuity index for drag versus point-by-point ablation with the use of proactive esophageal cooling

Westby G. Fisher, Tiffany Sharkoski, Erik Kulstad, William Zagrodzky et al.
European Heart Journal
Cardiac Arrhythmias and Treatments
article

Comparison of continuity index for drag versus point-by-point ablation with the use of proactive esophageal cooling

Westby G. Fisher, Tiffany Sharkoski, Erik Kulstad, William Zagrodzky, Mark D. Metzl, Natalie Putzel, Cathy L. Lazarus, A Ro, Jose Nazari, J.H. Sherman
article en

Abstract

Abstract Background Proactive esophageal cooling was recently FDA cleared to reduce the likelihood of esophageal injury during radiofrequency (RF) ablation for treatment of atrial fibrillation (AF). Long-term follow-up data have also shown improved freedom from arrhythmia with proactive esophageal cooling compared to luminal esophageal temperature (LET) monitoring during pulmonary vein isolation (PVI). One hypothesized mechanism is the greater lesion continuity attainable with the use of cooling, which can be quantified by the Continuity Index (which increases as continuity decreases). Increasing the Continuity Index increases gap formation, reduces isolation, and leads to greater AF recurrence; however, the difference in Continuity Index between ablating in a point-by-point fashion versus a dragging fashion has not previously been analyzed. Purpose To prospectively measure the Continuity Index of PVI cases and compare the dragging technique to point-by-point technique when using proactive esophageal cooling or LET monitoring. Methods We determined the Continuity Index for 101 patients: 77 prospectively utilizing proactive esophageal cooling, 24 retrospectively recorded using LET monitoring. Cases were completed by four operators, three utilizing point-by-point technique, and one using the drag technique. The Continuity Index was calculated by incrementing one unit for each lesion that did not border the previous lesion during the PVI (simplifying the original description of the Continuity Index). Results For cases using esophageal cooling and performing RF ablation using the drag technique, the average Continuity Index was 8.5 (range: 1 to 14). For point-by-point operators using esophageal cooling, the average Continuity Index was 1.2 (range: 0 to 9), representing a difference of 7.3 between the dragging technique and point-by-point ablation (p < .001). Patients receiving LET monitoring and RF ablation using the drag technique had Continuity Indices of 23.8 (range 15 to 30). Patients receiving LET monitoring and RF ablation using point-by-point ablation had Continuity Indices of 28.4 (range 16 to 50). Figure 1 compares these findings. Only 4 patients (5%) in the esophageal cooling group had a Continuity Index above 6 on either side (all in the dragging group), versus 100% (24) of patients in the LET monitoring group. Conclusion Proactive esophageal cooling during PVI is associated with a high degree of lesion continuity, as represented by the low Continuity Index. The drag technique resulted in a higher index than the point-by-point technique, but both provided significantly lower Continuity Indices (greater lesion continuity) than cases using LET monitoring. This may offer a mechanism for the greater freedom from arrhythmia seen to date with proactive cooling in long-term follow-up.Figure 1.Comparison of Continuity Index

European Heart JournalVol. 45(Supplement_1)
University of Southern California (US), Colorado College (US), Washington University in St. Louis (US), Evanston Hospital (US), University of Pennsylvania (US), The University of Texas Southwestern Medical Center (US)
Openalex Percentile: Top 65%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.