Quality evaluation of field safety notices of medical devices, across EU and non-EU countries

Abstract Introduction The EU Medical Device Regulation 2017/745 (MDR; article 87) requires device manufacturers to report to the relevant authorities any serious incident and any corrective action undertaken, through Field Safety Notices (FSN) whose content must be consistent in all Member States. Before MDR, due to the lack of harmonized global standards for reporting, FSNs were published independently by each country with different formats, styles, nomenclatures and languages. This heterogeneity makes it challenging to use such historical data for trend analysis in post-market surveillance (PMS). Purpose 1) To assess the quality of the FSNs issued by EU or non-EU national authorities for analysing historical trends. 2) To group countries based on such quality, and test for differences between EU and non-EU countries. Methods As part of the CORE-MD project coordinated by the ESC, all FSN information publicly available as HTML text (excluding linked PDFs) was retrieved automatically from national authorities’ websites, using the CORE-MD PMS tool [1]. For each FSN, a score was assigned to each field of interest (1 or 2, according to importance for building trends) (see Figure 1), and a Percentage Score (PS) computed as the % of the ratio between the scores’ sum and the maximum possible score (10). Based on PS, FSNs were categorized as Excellent (≥90), Very Good (80≤PS<90), Good (70≤PS<80), Medium (60≤PS<70), or Unqualified (<60). For each country, the distribution of FSN categories was computed. Hierarchical Agglomerative Clustering (HAC) was used to group countries based on their similarities, and differences between EU and non-EU countries were assessed by Mann-Whitney U test. Results 126,405 FSNs published before 31/12/2022 were retrieved and analyzed. The % of FSN in which each field was available is shown by country in Figure 1. Manufacturer and Device were clearly described, but more detailed device information was often absent. The Description, with the reason for publishing the FSN, and the Device Category, allowing nomenclature categorization, were provided only by a few countries. The % of FSNs in each quality category is reported in Figure 2: only Italy provided some FSNs (31%) classified as Excellent. The HAC analysis identified three clusters: "high-quality FSNs" (Czechia, Denmark, Italy, Sweden, the Netherlands, and the USA), "moderate-quality FSNs" (Australia, Canada, Greece, Ireland, Latvia, Slovenia, Spain, and the UK), and "low-quality FSNs" (Croatia, Estonia, France, Germany, Poland, and Portugal). There was no difference between EU and non-EU countries. Conclusions A significant discrepancy was observed in the quality of FSNs retrieved from different countries, highlighting the difficulty in using such data to analyse trends, for example in reports of cardiovascular devices. This study reinforces the need for a global minimum reporting standard, to facilitate more effective PMS.Figure 1Figure 2

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.3420
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality evaluation of field safety notices of medical devices, across EU and non-EU countries

Enrico G. Caiani, Yong Ren, Alan G. Fraser
European Heart Journal
Quality and Safety in Healthcare
article

Quality evaluation of field safety notices of medical devices, across EU and non-EU countries

Enrico G. Caiani, Yong Ren, Alan G. Fraser
article en

Abstract

Abstract Introduction The EU Medical Device Regulation 2017/745 (MDR; article 87) requires device manufacturers to report to the relevant authorities any serious incident and any corrective action undertaken, through Field Safety Notices (FSN) whose content must be consistent in all Member States. Before MDR, due to the lack of harmonized global standards for reporting, FSNs were published independently by each country with different formats, styles, nomenclatures and languages. This heterogeneity makes it challenging to use such historical data for trend analysis in post-market surveillance (PMS). Purpose 1) To assess the quality of the FSNs issued by EU or non-EU national authorities for analysing historical trends. 2) To group countries based on such quality, and test for differences between EU and non-EU countries. Methods As part of the CORE-MD project coordinated by the ESC, all FSN information publicly available as HTML text (excluding linked PDFs) was retrieved automatically from national authorities’ websites, using the CORE-MD PMS tool [1]. For each FSN, a score was assigned to each field of interest (1 or 2, according to importance for building trends) (see Figure 1), and a Percentage Score (PS) computed as the % of the ratio between the scores’ sum and the maximum possible score (10). Based on PS, FSNs were categorized as Excellent (≥90), Very Good (80≤PS<90), Good (70≤PS<80), Medium (60≤PS<70), or Unqualified (<60). For each country, the distribution of FSN categories was computed. Hierarchical Agglomerative Clustering (HAC) was used to group countries based on their similarities, and differences between EU and non-EU countries were assessed by Mann-Whitney U test. Results 126,405 FSNs published before 31/12/2022 were retrieved and analyzed. The % of FSN in which each field was available is shown by country in Figure 1. Manufacturer and Device were clearly described, but more detailed device information was often absent. The Description, with the reason for publishing the FSN, and the Device Category, allowing nomenclature categorization, were provided only by a few countries. The % of FSNs in each quality category is reported in Figure 2: only Italy provided some FSNs (31%) classified as Excellent. The HAC analysis identified three clusters: "high-quality FSNs" (Czechia, Denmark, Italy, Sweden, the Netherlands, and the USA), "moderate-quality FSNs" (Australia, Canada, Greece, Ireland, Latvia, Slovenia, Spain, and the UK), and "low-quality FSNs" (Croatia, Estonia, France, Germany, Poland, and Portugal). There was no difference between EU and non-EU countries. Conclusions A significant discrepancy was observed in the quality of FSNs retrieved from different countries, highlighting the difficulty in using such data to analyse trends, for example in reports of cardiovascular devices. This study reinforces the need for a global minimum reporting standard, to facilitate more effective PMS.Figure 1Figure 2

European Heart JournalVol. 45(Supplement_1)
Cardiff University (GB), Politecnico di Milano (IT)
European Commission
Openalex Percentile: Top 62%
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.