Atrial fibrillation ablation with temperature-controlled very high-power short-duration, with conventional-power, or with combined modes in clinical practice: 12-month outcomes from SECURE

Abstract Background/Introduction QDM is a contact force-sensing catheter optimized for temperature-controlled radiofrequency ablation in very high-power short-duration (vHPSD; 90 W/4 s), conventional-power (CP; ≤50 W), or vHPSD/CP combined modes. Purpose Evaluate 12-month (M) safety and effectiveness of atrial fibrillation (AF) ablations performed with 3 different ablation modes using the QDM catheter, a frequently used single tip catheter in Europe. Methods SECURE is an observational, prospective, post-market study collecting data on procedures performed per standard-of-care with up to 12M follow-up, in 16 enrolling sites in 8 countries across Europe and Israel. Index pulmonary vein (PV) isolation procedures are reported here. The primary safety endpoint is the incidence of primary adverse events (PAEs); the primary effectiveness endpoint is PV isolation (entrance block) of targeted PVs. Additional endpoints include acute reconnections, incidence of char and steam pop, 12M recurrence, repeat ablation, and healthcare utilization (HCU). Ablation strategy, impedance drop, and catheter stability were analysed with a cloud-based data management and artificial intelligence-powered algorithm. Results Since Feb 2021, 291 patients (70% male, 61±11 yrs, CHA2DS2-VASc 1.9±1.5) completed 12M follow-up in 11 sites. Most patients (175 [60.1%]) were treated for paroxysmal (PAF), 85 (29.2%) for persistent (PsAF), and 31 (10.7%) for long-standing persistent (LsPsAF) AF, under general anaesthesia (44.0%) or sedation (55.7%). A combined workflow was used in 66.0% of ablations, while vHPSD-only and CP-only were used in 13.4% and 20.6% of cases, respectively. Four (1.4%) patients reported a PAE; all resolved, and none were catheter related. Acute effectiveness was achieved in all targeted veins; 5.0% of PVs had acute reconnections, with the most common being the left superior PV. No steam pops were reported; 2 counts of char were noted in PsAF ablations with combined modes. Freedom from documented recurrence were 95.8% for PAF, 88.1% for PsAF, and 73.5% for LsPsAF (Fig 1). Freedom from repeat ablation post-blanking and freedom from cardioversion were both 96.8%. Cardiovascular hospitalization and emergency room visits decreased by 65.3% and 76.5% from 12M pre- to 12M post-procedure, respectively. Application-level analyses showed larger impedance drop and greater catheter stability with vHPSD. In procedures using a combined workflow, vHPSD was used preferentially for most PV segments, except anteriorly on the left wide area circumferential ablation (Fig 2). Conclusions Real-world data demonstrates that AF ablation with the QDM catheter is safe and effective with a low PAE rate, high 12M effectiveness, and reduced HCU. Most procedures were done with combined modes, where vHPSD was used preferentially, except at left anterior thicker regions. Larger impedance drop and greater catheter stability were observed with vHPSD applications.

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.553
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Atrial fibrillation ablation with temperature-controlled very high-power short-duration, with conventional-power, or with combined modes in clinical practice: 12-month outcomes from SECURE

Isabel Deisenhofer, Helmut Pürerfellner, Thomas Phlips, G. André Ng et al.
European Heart Journal
Atrial Fibrillation Management and Outcomes
article

Atrial fibrillation ablation with temperature-controlled very high-power short-duration, with conventional-power, or with combined modes in clinical practice: 12-month outcomes from SECURE

Isabel Deisenhofer, Helmut Pürerfellner, Thomas Phlips, G. André Ng, Pier D. Lambiase, Johan Vijgen, Raphaël Rosso, Andrea Sarkozy, Federico Quadrini, Mattias Duytschaever, Tom De Potter, Thomas Deneke, Maria Grimaldi
article en

Abstract

Abstract Background/Introduction QDM is a contact force-sensing catheter optimized for temperature-controlled radiofrequency ablation in very high-power short-duration (vHPSD; 90 W/4 s), conventional-power (CP; ≤50 W), or vHPSD/CP combined modes. Purpose Evaluate 12-month (M) safety and effectiveness of atrial fibrillation (AF) ablations performed with 3 different ablation modes using the QDM catheter, a frequently used single tip catheter in Europe. Methods SECURE is an observational, prospective, post-market study collecting data on procedures performed per standard-of-care with up to 12M follow-up, in 16 enrolling sites in 8 countries across Europe and Israel. Index pulmonary vein (PV) isolation procedures are reported here. The primary safety endpoint is the incidence of primary adverse events (PAEs); the primary effectiveness endpoint is PV isolation (entrance block) of targeted PVs. Additional endpoints include acute reconnections, incidence of char and steam pop, 12M recurrence, repeat ablation, and healthcare utilization (HCU). Ablation strategy, impedance drop, and catheter stability were analysed with a cloud-based data management and artificial intelligence-powered algorithm. Results Since Feb 2021, 291 patients (70% male, 61±11 yrs, CHA2DS2-VASc 1.9±1.5) completed 12M follow-up in 11 sites. Most patients (175 [60.1%]) were treated for paroxysmal (PAF), 85 (29.2%) for persistent (PsAF), and 31 (10.7%) for long-standing persistent (LsPsAF) AF, under general anaesthesia (44.0%) or sedation (55.7%). A combined workflow was used in 66.0% of ablations, while vHPSD-only and CP-only were used in 13.4% and 20.6% of cases, respectively. Four (1.4%) patients reported a PAE; all resolved, and none were catheter related. Acute effectiveness was achieved in all targeted veins; 5.0% of PVs had acute reconnections, with the most common being the left superior PV. No steam pops were reported; 2 counts of char were noted in PsAF ablations with combined modes. Freedom from documented recurrence were 95.8% for PAF, 88.1% for PsAF, and 73.5% for LsPsAF (Fig 1). Freedom from repeat ablation post-blanking and freedom from cardioversion were both 96.8%. Cardiovascular hospitalization and emergency room visits decreased by 65.3% and 76.5% from 12M pre- to 12M post-procedure, respectively. Application-level analyses showed larger impedance drop and greater catheter stability with vHPSD. In procedures using a combined workflow, vHPSD was used preferentially for most PV segments, except anteriorly on the left wide area circumferential ablation (Fig 2). Conclusions Real-world data demonstrates that AF ablation with the QDM catheter is safe and effective with a low PAE rate, high 12M effectiveness, and reduced HCU. Most procedures were done with combined modes, where vHPSD was used preferentially, except at left anterior thicker regions. Larger impedance drop and greater catheter stability were observed with vHPSD applications.

European Heart JournalVol. 45(Supplement_1)
Rhön-Klinikum (DE), University of Leicester (GB), Tel Aviv Sourasky Medical Center (IL), Glenfield Hospital (GB), British Heart Foundation (GB), Jessa Hospital (BE), Onze Lieve Vrouwziekenhuis Hospital (BE), Antwerp University Hospital (BE), Krankenhaus der Elisabethinen (AT), Ospedale Generale Regionale Francesco Miulli (IT), Deutsches Herzzentrum München (DE), University College London (GB), Technical University of Munich (DE)
Openalex Percentile: Top 66%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.