Pacing sensation with a second generation intercostal extravascular ICD lead

Abstract Background/Introduction An investigational intercostal extravascular ICD lead (EV-ICD Lead) has been developed to provide antitachycardia pacing and shock treatment of ventricular tachyarrhythmias when connected to a commercially available ICD pulse generator. Unlike other available extravascular ICD systems that achieve pacing with ring electrodes which transmit undirected current to cardiac and non-cardiac tissue alike, the intercostal EV-ICD Lead uses button-shaped electrodes backed by an insulative paddle to direct current toward excitable myocytes while minimizing undesired excitation of the adjacent tissues. Purpose The purpose of this analysis was to evaluate pacing sensation with the intercostal EV-ICD Lead connected to a commercially-available ICD pulse generator programmed to the maximum voltage output and pulse width settings. Methods In the STEP ICD Study (n=20), pacing sensation was evaluated at the pre-discharge visit while pacing at the maximum programmable voltage output (7.5 or 8.0 V, depending on manufacturer) and pulse width (1.5 or 2.0 ms). Sensation was reported by the patient as one of four choices: No Sensation, Sensation Only, Tolerable Pain or Intolerable Pain. Results Pacing sensation was assessed in 18 of 20 patients. One (1) patient did not undergo a pacing evaluation due to atrial fibrillation with rapid ventricular response. Responses were as follows: No Sensation in 4 (22%) patients, Sensation Only in 10 (56%) patients, Tolerable Pain in 3 (17%) and Intolerable Pain in 1 (6%). In no patient was the pain so severe that pacing capture threshold testing could not be completed. Of note, the single report of intolerable pain was only reported in the sitting posture. When tested supine, the patient reported sensation only. Conclusion(s) The design of the investigational intercostal EV-ICD Lead, with button-shaped electrodes backed by an insulative paddle, produced low levels of patient sensation at maximum output and pulse width. These sensation levels represent an improvement over designs that employ ring electrodes and indicate that antitachycardia pacing will be well tolerated with the novel EV-ICD Lead. These results corroborate the predictions about patient discomfort from in silico models.

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.715
Citations
1
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pacing sensation with a second generation intercostal extravascular ICD lead

Shawn Chen, Adrián Ebner, M C Burke, Michael Husby et al.
1 citations
European Heart Journal
Cardiac pacing and defibrillation studies
0.40
article

Pacing sensation with a second generation intercostal extravascular ICD lead

Shawn Chen, Adrián Ebner, M C Burke, Michael Husby, RE Knops, Alan Marcovecchio, D E Scheck, Marie O’Connor, Ravina Sanghera, Sagar N. Doshi
article en
1 citations

Abstract

Abstract Background/Introduction An investigational intercostal extravascular ICD lead (EV-ICD Lead) has been developed to provide antitachycardia pacing and shock treatment of ventricular tachyarrhythmias when connected to a commercially available ICD pulse generator. Unlike other available extravascular ICD systems that achieve pacing with ring electrodes which transmit undirected current to cardiac and non-cardiac tissue alike, the intercostal EV-ICD Lead uses button-shaped electrodes backed by an insulative paddle to direct current toward excitable myocytes while minimizing undesired excitation of the adjacent tissues. Purpose The purpose of this analysis was to evaluate pacing sensation with the intercostal EV-ICD Lead connected to a commercially-available ICD pulse generator programmed to the maximum voltage output and pulse width settings. Methods In the STEP ICD Study (n=20), pacing sensation was evaluated at the pre-discharge visit while pacing at the maximum programmable voltage output (7.5 or 8.0 V, depending on manufacturer) and pulse width (1.5 or 2.0 ms). Sensation was reported by the patient as one of four choices: No Sensation, Sensation Only, Tolerable Pain or Intolerable Pain. Results Pacing sensation was assessed in 18 of 20 patients. One (1) patient did not undergo a pacing evaluation due to atrial fibrillation with rapid ventricular response. Responses were as follows: No Sensation in 4 (22%) patients, Sensation Only in 10 (56%) patients, Tolerable Pain in 3 (17%) and Intolerable Pain in 1 (6%). In no patient was the pain so severe that pacing capture threshold testing could not be completed. Of note, the single report of intolerable pain was only reported in the sitting posture. When tested supine, the patient reported sensation only. Conclusion(s) The design of the investigational intercostal EV-ICD Lead, with button-shaped electrodes backed by an insulative paddle, produced low levels of patient sensation at maximum output and pulse width. These sensation levels represent an improvement over designs that employ ring electrodes and indicate that antitachycardia pacing will be well tolerated with the novel EV-ICD Lead. These results corroborate the predictions about patient discomfort from in silico models.

European Heart JournalVol. 45(Supplement_1)
CorVita Science Foundation (US), Pacific Heart Institute (US), Amsterdam University Medical Centers (NL)
Openalex Percentile: Top 31%
Cardiac pacing and defibrillation studies
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.